Pro rychlou technickou pomoc nás mÛÏete kontaktovat na tûchto adresách
UÏivatelská pfiíruãka
Severní Amerika USA: One Omega Drive, BOX 4047, Stamford, CT 06907-0047 Tel: (203) 359-1660 Fax: (203) 359-7700 e-mail:
[email protected]
SERIES
THE NEW STANDARD FOR QUALITY
Canada: 976 Bergar, Laval (Quebec) H7L 5A1 Tel: (514)-856-6928 Fax: (514) 856-6886 e-mail:
[email protected]
INFCP
DP25-E
Procesní indikátor série INFINITY
Mexico: Tel: (95) 800-826-6342 Fax: (95) 203-359-7807 e-mail:
[email protected] âeská Republika: Rudé armády 1868, 733 01 Karviná 8 Tel: 420-69-6311899 Fax: 420-69-6311114 e-mail:
[email protected] [email protected]
®
http://www.omega.com e-mail:
[email protected]
Anglie: 25 Swannington Road, Broughton Astley, Tel: (514) Leicestershire LE9 6TU, Anglie Tel:44 (1455) 285520 nebo 0800-488-488 Fax:44 (1455) 283912 e-mail:
[email protected] Nûmecko: Daimlerstrasse 26, D-75392 Deckenpfronn Tel: 49 (07056) 3017 nebo 0130 11 21 66
[email protected] Francie: 9 Rue Denis Papin, 78190 Trappes Tel: 33 0130-621-400 Fax: 33 0130-699-120 e-mail:
[email protected] http://www.newport.cz
OMEGAnetSM ON-LINE SERVICE
e-mail:
[email protected]
http://www.omega.com e-mail:
[email protected]
11540ML-03 Rev. A
Obsah 1.0 2.0 3.0 4.0 5.0 6.1 6.2 6.3 7.0 7.1 8.0 9.0 10.0 11.0
12.0 12.1 12.2 12.3 13.0 14.0
MontáÏ Bezpeãnostní upozornûní Zapojení pinÛ pfiipojovací svorkovnice Hardware - konfigurace napájení snímaãe a typu vstupního signálu Aktivace programov˘ch závor Pfiipojení 3-vodiãového snímaãe s napûÈov˘m signálem Pfiipojení 4-vodiãového snímaãe s napûÈov˘m signálem Pfiipojení snímaãe s proudov˘m v˘stupem 0/4 - 20 mA Pfiipojení napájecího napûtí u pfiístrojÛ 230-115 Vst Pfiipojení napájecího napûtí u pfiístrojÛ 9-32 Vss Pfiipojení analogového v˘stupu Pfiipojení reléov˘ch v˘stupÛ Nastavení mezních hodnot Programování INTP (INPUT) Nastavení typu vstupního signálu DEC.P Nastavení pozice desetinné teãky RD.S.O Nastavení zobrazení RD.CF Aktivace funkce tarování, nastavení interního rozli‰ení S1.CF Parametry mezní hodnoty 1 S2.CF Parametry mezní hodnoty 2 S1.DB Nastavení hysterese mezní hodnoty 1 S2.DB Nastavení hysterese mezní hodnoty 2 OT.FC Konfigurace analogového v˘stupu OT.S.O Pfiifiazení indikace k analogovému v˘stupu P.BAND Proporcionální pásmo pfii P-regulaãní funkci M.RST Komp.odchylky od Ïádané hodnoty pfii P-regulaãní funkci analogového v˘stupu LK.CF Programové závory pro mezní hodnoty a reset Funkce klávesnice v RUN-Modu Vyvolání uchováné MAX-hodnoty Tarování Kvitace pfii pfiekroãení mezní hodnoty Hlá‰ení bûhem programování a chybová hlá‰ení Specifikace
Kde mohu nalézt v‰echno co potfiebuji pro mûfiení a regulaci ? u OMEGY .... samozfiejmû ! TEPLOTA:
SBùR DAT:
Termoãlánky, Pt100, termistory, konektory, panely a pfiíslu‰enství Vodiãe: termoãlánky, Pt100, termistory Kalibrátory a referenãní nulov˘ bod Zapisovaãe, regulátory a monitory procesu Infrapfiístroje
Software pro sbûr dat a inÏen˘rské aplikace Zafiízení pro sbûr a zpracován dat Zásuvné karty do PC a Apple poãítaãÛ Datalogry Zapisovaãe, tiskárny a plotry
TLAK, TAH A SÍLA
TOPNÁ TùLESA
Tlakové snímaãe a tenzometry VáÏní ãidla Snímaãe posunutí Pfiístroje a pfiísllu‰enství
PRÒTOK A HLADINA
Rotametry, hmotnostní prÛtokomûry Mûfiení rychlosti proudûní vzduchu Turbínkové a lopatkové prÛtokomûry âítaãe, totalizéry a dávkovaãe
pH a VODIVOST
pH elektrody, testery a pfiíslu‰enství Laboratorní a pfienosné pH metry a konduktometry Regulátory, kalibrátory, simulátory a pumpy PrÛmyslové mûfiiãe pH a vodivosti
Topné kabely Válcová a prouÏková topná tûlesa Ponorné ohfiívaãe Ohebné topné pásky Laboratorní ohfiívaãe
OCHRANA ÎIVOTNÍHO PROST¤EDÍ
Mûfiící a fiídící pfiístroje Refraktometry Pumpy a hadiãky Monitorování pÛdy, vzduchu a vody Testování uÏitkov˘ch a odpadních vod pfiístroje pro mûfiení pH, vodivosti a rozpu‰tûného kyslíku
Tento dokument nesmí b˘t kopírován, reprodukován, pfiekládán, nebo redukován na jiné elektronické médium nebo do ti‰tûné formy, jako celek nebo ãást bez pfiedchozího písemného svolení firmy OMEGA ENGINEERING, INC.
V‰eobecn˘ popis Model INFCP je moderní 4-místn˘ mikroprocesorov˘ mûfiící pfiístroj pro procesní signály Nastavitelné mûfiící rozsahy: 0-100 mV, 650 mV, 0-10 V, 65V, 0-20 mA a 4-20 mA.
© Copyright 1996 OMEGA, ENGINEERING INC. V·ECHNA PRÁVA VYHRAZENA
V základním provedení jej lze pouÏívat jako indikátor procesních veliãin, umoÏÀuje napájet pfiipojen˘ snímaã. Základní verzi lze roz‰ífiit o 2 releové v˘stupy a analogov˘ v˘stup (toto je nutno specifikovat v dobû objednávky). Programování pfiístroje se provádí jednodu‰e z ãelního panelu. V˘robek je chránûn jedním nebo více z následujících patentÛ : U.S. Patents Des. 336,895; 5,274,577 / FRANCE Brevet No. 91 12756 / SPAIN 2039150 / U.K. Patent No. 2248954, 22498371 / CANADA 2052600 / ITALY 1249456, 1250938 / GERMANY DE 4134398C2 a dal‰ími mezinárodními patenty.
Standartní funkce: - indikaãní rozsah - 1999...9999, pfiesnost 60,03%, digitální filtr - rozsah mûfiení 0-100 mV, 650mV, 0-10V, 65V, 0-20 mA a 4-20 mA
-2-
Záruka a odpovûdnost Potvrzení objednávek: Pokud podáte objednávku telefonem, informujte prodejce, Ïe písemné potvrzení objednávky bude následovat. Aby se pfiede‰lo duplikacím, oznaãte vaãe potvrzení poznámkou 'Pouze potvrzení, neobjednávejte znovu' a pfiipojte uveìte jméno prodejce. Písemné objednávky: Písemné objednávky vítáme. Znáte-li dobfie v˘robky OMEGA a nepotfiebujete kozultovat prodejce, posílejte nebo faxujte písemné objednávky. Pro rychlé a úspû‰né vyfiízení Va‰í objednávky prosím uveìte: Va‰e ãíslo objednávky âíslo úãtu a adresu, kam se má objednávka zaslat Kód v˘robku a popis objednan˘ch poloÏek Telefonní ãíslo Ïadatele Termíny a podmínky: OMEGA prodluÏuje platební termíny na ãist˘ch 30 dní pro v‰echny zákazníky, ktefií u ní mají otevfien˘ úãet. OMEGA uvítá nové úãty a zpracovávat objednávky placené pfiedem nebo kreditní kartou, jakmile je otevfien˘ úãet ustaven. Zmûny objednávky: ZpÛsob doruãení, zmûny objednávky a zru‰ení objednávky prosím projednejte s expediãním oddûlením fy OMEGA. Zvlá‰tní podmínky: Má-li b˘t zafiízení pouÏito v radioaktivním prostfiedí, zákazník musí zajistit, aby firma OMEGA neutrpûla Ïádnou ‰kodu a nenesla Ïádnou odpovûdnost, aÈ pfii tomto pouÏití dojde k ãemukoliv. Zafiízení prodávaná firmou OMEGA nejsou urãeny pro aplikace v lékafiství nebo uÏití na lidech. OMEGA nepfiijímá ve smyslu v˘‰euveden˘ch základních záruãních podmínek Ïádnou odpovûdnost, jsou-li její v˘robky pouÏívány v lékafiství nebo na lidech, nebo jsou-li zneuÏívány jak˘mkoli zpÛsobem. Ceny: ZboÏí je prodáváno za ceny platné v dobû prodeje. Ceny uvedené v tomto katalogu mohou b˘t zastaralé a mohou se bez upozornûní mûnit. Aktuální ceny Vám sdûlí prodejní oddûlení. OMEGA Vám ochotnû poskytne tyto informace po‰tou, faxem nebo telefonicky. MnoÏstevní slevy: U mnoha poloÏek existuje tabulka mnoÏstevních slev. Pro informace o velk˘ch mnoÏstvích a produktech, u kter˘ch není zpracována tabulka mnoÏstevních slev, se obraÈte na prodejní oddûlení. Kreditní karty: OMEGA umoÏÀuje platbu nejpouÏívanûj‰ími platebními kartami VISA a Acces/MasterCard. Minimální platba: Minimální platba je 500 Kã. Dopravné: V cenách zboÏí jsou náklady na po‰tovné vãetnû poji‰tûní. Jiné zpÛsoby dopravy jsou moÏné na zvlá‰tní Ïádost.
-
napájení snímaãe 24V-25mA, 12V-50mA, 10V-120mA, 5V-60mA DIN - rozmûr 96 x 48 mm MAX - pamûÈ maximální hodnoty Funkce tarování 2 mezní hodnoty s reléov˘m v˘stupem max. 230 VAC-30VDC, 6A ‰kálovan˘ analogov˘ v˘stup 0-10V, 0-4 - 20 mA moÏná P-regulaãní funkce analogového v˘stupu nejjednodu‰‰í programování, vícestupÀové programové blokování programování z ãelního panelu 230 VAC, 115 VAC a 9-32 VDC volitelné napájecí napûtí ãelní krytí IP 65 moÏno doobjednat
1.0 MontáÏ âelní rozmûry pfiístroje jsou 96 x 48 mm, rozmûry v˘fiezu v panelu 92 x 45 mm. ZasuÀte pfiístroj zepfiedu pfies v˘fiez a upevnûte jej pomocí pfiiloÏen˘ch upevÀovacích dílÛ.
2.0 Bezpeãnostní upozornûní Pfiístrojová fiada INCFP odpovídá stupni ochrany IEC 348, tfiída II a VDE 0411. Dbejte, prosím, pfii pfiipojení snímaãe na následující souvislosti: Vstup mûfieného signálu a analogov˘ v˘stup nejsou galvanicky oddûleny. Pfii pouÏívání pfiístrojÛ fiady INFINITY dbejte v‰ech bezpeãnostních opatfiení uveden˘ch v platn˘ch pfiedpisech pro práci na elektrick˘ch zafiízeních dle norem âSN. Odpojte pfied vynûtím INFCP z krytu v‰echna pfiipojovací vedení. StupeÀ ochrany (krytí): âelní stûna IP 54 podle DIN 40050. Zadní stûna IP 20 podle DIN 40050
ZadrÏení právního nároku: V‰echny v˘robky prodávané firmou OMEGA zÛstávají jejím vlastnictvím, dokud nejsou úplnû zaplaceny. Poznámka k záruãním podmínkám: Záruku nelze pfiedat nebo postoupit tfietí stranû. Je omezena pouze na kupujícího. V‰echny záruky jsou povaÏovány za neplatné, dojde-li k pfievodu na tfietí stranu, pokud není zámûr pfiedat v˘robek tfietí stranû v˘raznû uveden na objednávce nebo pokud zákazník není registrovan˘m prodejcem v˘robkÛ OMEGA. PoÏadavky na v˘mûnu nebo opravu: Prosím, zasílejte v‰echny poÏadavky na záruãní i pozáruãní opravy pfiímo na servisní oddûlení. Pfied zasláním pfiístroje kontaktujte servisní oddûlení, které Vám pfiidûlí autorizované v˘mûnné ãíslo (AR) a adresu, kam v˘robek zaslat. Pfiidûlené ãíslo AR by mûlo b˘t uvedeno na vnûj‰í stranû zabaleného v˘robku. Aby nedocházelo ke zpoÏdûní, ujistûte se prosím, Ïe zásilka obsahuje: âíslo objednávky a faktury: Va‰e jméno, adresu a telefonní ãíslo âíslo modelu a série. Instrukce pro opravu. OEM oprávnûní: KaÏd˘ kvalifikovan˘ v˘robce mÛÏe získat OEM oprávnûní. Formuláfi Ïádosti Vám pfiedá OEM prodejní oddûlení. Vzhled a vlastnosti: OMEGA si vyhrazuje právo zmûnit vzhled a vlastnosti jakéhokoliv v˘robku pfii uplatnûní technick˘ch zlep‰ení nebo jin˘ch nezbytn˘ch poÏadavkÛ.
3.0 Zapojení pinÛ pfiipojovací svorkovnice Pin
Funkce
TB 1-1 -2 -3 -4 -5 -6 -7 -8 -9 10 11 12 TB 2-1 -2
Mezní hodnota 1 spínací kontakt, 230 VAC/30 VDC, 6A Mezní hodnota 1 rozepínací konatkt, 230 VAC/30 VDC, 6A Mezní hodnota 1 spoleãn˘ kontakt Mezní hodnota 2 spínací kontakt 230 VAC/30 VDC, 6A Mezní hodnota 2 rozepínací kontakt 230 VAC/30 VDC, 6A Mezní hodnota 2 spoleãn˘ kontakt Fáze st-napájení pfii ss napájení není obsazen Nulov˘ vodiã st-napájení +ss napájení Ochrann˘ vodiã st-napájení - ss napájení + Analogov˘ v˘stup 0-10 V + Analogov˘ v˘stup 0/4-20 mA Analogov˘ v˘stup GND - napájení snímaãe 5V, 10V a 12V + napájení snímaãe 5V, 10V a 12V -3-
-3 -5 -6 -7
+ + + -
Poznámky:
vstupní signál 0/4 - 20 mA napájení snímaãe 24V vstupní signál pfii napûÈov˘ch signálech signální vstup a GND pro 24V napájení snímaãe
J 1 (1-2) vstup tarování, nastavení prostfiednictvím spínacího kontaktu
4.0 Hardware - nastavení napájení snímaãe a typu vstupního signálu DIP-spínaã S1 se nachází na zadní stranû pfiístroje. Pomocí DIP-spínaãe S1 bude nastaveno napájení snímaãe a typ mûfieného signálu. Nastavení se mÛÏe dosáhnout bez rozdûlání pfiístroje. C = Closed (spínaã dole) O = Open (spínaã nahofie) Funkce
1
2
3
4
5
6
7
8
Napájení snímaãe interní 5/10/12 V Napájení snímaãe externí Napájení snímaãe 12 V Napájení snímaãe 10 V Napájení snímaãe 5 V
C 0 C C C
-
-
-
-
0 C C
0 0 C
-
24 V napájení snímaãe je k dispozici nezávisle na nastavení na 5/10/12 V. 24 V napájení snímaãe je vÏdy na pinech TB 2 Pin 5(+) a TB 2 Pin 7 (-). 5/10/12 V napájení snímaãe je vÏdy na pinech TB 2 Pin 2 (+E) a TB 2 Pin 1 (-E). Mûfien˘ signál 0 - 100 mV Mûfien˘ signál 650 mV Mûfien˘ signál 65 V Mûfien˘ signál Mûfien˘ signál
-
0 0 C C 0
C C 0 0 C
0 0 0 0 C
0 C C 0 0
-
-
0 C C C 0
5.0 Aktivace programov˘ch závor Pomocí zástrãky S3 bude aktivována programová závora. Funkce zástrãky S3A instalována: v‰echna ãelní tlaãítka jsou funkãní S3A neinstalována: v‰echna ãelní tlaãítka jsou nefunkãní S3B instalována: pouze pro kalibraci pfiístroje S3B neinstalována: provozní mód S3C instalována: analogov˘ v˘stup bez pfiedfiadného odporu S3C neinstalována: analogov˘ v˘stup s pfiedfiadn˘m odporem S3-C smí b˘t instalována jen tehdy, kdyÏ je v˘stup pfiipojen na pfiístroje s nízk˘m vstupním odporem. Dbejte, prosím, na to, Ïe pfii instalováni S3-C není analogov˘ v˘stup ji‰tûn proti zkratu. S3D není obsazen (pro budoucí pouÏití) S3-E je -li instalován bez S3-B, dojde k vyfiazení tlaãítka MENU. Pfii stlaãení tlaãítka MENU se objeví na displeji “LOCK“.
-4-
Displej
6.1 Pfiipojení 3-vodiãvého snímaãe s napûÈov˘m signálem
14 mm, 14-segmentová ãervená LED-indikace
Interní napájení snímaãe
Externí napájení snímaãe
Analogovû/digitální pfievodník Technika Interní rozli‰ení Mûfi. pfiírÛstek Indikace polarity
Dual Slope 15 bit 3 mûfiení/sek automaticky
Pfiesnost pfii 25°C Nejvût‰í chyba Teplotní koeficient Doba odezvy Teplotní prÛbûh
¨NAP
¨ZEM
60,03 % indikace 6Digit 650 ppm/°C 1 - 2 sek. 30 min.
6.2 Pfiipojení 4-vodiãvého snímaãe s napûÈov˘m signálem
Analogov˘ v˘stup Signální úroveÀ Funkce Linearita Regulace zatíÏení Regulace síÈ. napûtí Skoková odezva
SIG
0-10 V, 4-20 mA, 0-20 mA Pevné pfiifiazení zobrazení k analogovému v˘stupu nebo P-regulaãní funkce ve spojení s mezní hodnotou 1 0,2 % 1,1 % 0,02 % pro zmûny síÈ.napûtí o 1 V 2-3 sek. na 99% koneãné hodnoty
Interní napájení snímaãe
Externí napájení snímaãe
Napájení snímaãe NAPùTÍ
24V max. 25 mA, 12V max. 50 mA, 10V max. 120 mA, 5V max. 60 mA
Napájení Napûtí Kmitoãet V˘kon
115 VAC nebo 230 VAC 615%;9-32 VDC, galvanicky oddûleno 50 - 60 Hz 6 Watt
Interní napájení snímaãe
Provozní podmínky Provozní teplota Skladovací teplota Relativní vlhkost
6.3 Pfiipojení snímaãe s proudov˘m signálem 0/4 - 20 mA Externí napájení snímaãe
0 - 50°C -40...+85°C 90% pfii 40°C (nekondenzovaná)
Mechanická údaje V˘fiez v panelu Rozmûry Hmotnost Materiál krytu
92 x 45 mm 96 x 48 x 156 mm (‰ x v x H) 574 g Polykarbonát, 94 V-0 UL
Kontakty relé ZatíÏitelnost Provedení Odru‰ení
max. 230V/30 VDC, 6A pfiepínací kontakty RC-ãlen s 2500 pF, 200 Ohm - 16 -
-5-
Pfiipojení snímaãe s proudov˘m signálem 0/4 - 20 mA (zdroj proudu)
12.2 TAROVÁNI Hodnota parametru RD.CF musí b˘t nastaven na R1=T. Stisknûte tlaãítko TARE
Dojde k vynulování hodnoty na displeji na 0000
Pro opûtné zru‰ení funce tarování stisknûte tlaãítko RESET.
12.3 Kvitace pfii pfiekroãení mezní hodnoty Kvitace pfii pfiekroãení mezní hodnoty je nutná jen pfii programování S1.CF/S2.CF S.2=L (Potvrzovací reÏim pro mezní hodnoty). Stisknûte tlaãítko RESET
Na displeji se krátce objeví SP.RS , pfiekroãení mezní hodnoty bude kvitováno (potvrzeno operátorem)
13. Hlá‰ení bûhem programování a chybová hlá‰ení 7.0 Pfiipojení napájecího napûtí u pfiístrojÛ 230/115 VAC
L1 N PE
7.1 Pfiipojení napájecího napûtí pfii 9-32 VDC pfiístrojích
Hlá‰ení Funkce
Popis
LOCK
Zástrãka S3-E instalována, klávesnice aÏ na odpovûì resp. nastavení mezní hodnoty blokována. Naproti tomu S3-E odpojit. Hard-Reset pfii zapnutí resp. pfii ukonãení programování Typ vstupního signálu napfi. 0=100 mV, 0-20 mA.. Desetinná teãka ·kálování zobrazení Nastavení interního rozli‰ení a tarování Nastavení mezní hodnoty 1 Nastavení mezní hodnoty 2 Hysterese mezní hodnoty 1 Hysterese mezní hodnoty 2 Nastavení analogového v˘stupu Pfiekroãení rozsahu vstupu Pfiekroãení rozsahu vstupu Zvolen˘ signál nemÛÏe b˘t se zvolenou citlivostí zobrazen Pfiekroãení zobrazeného indikaãního rozsahu Podkroãení zobrazeného indikaãního rozsahu Chyba ve 2-soufiadnicovém ‰kálování Zpûtné nastavení tarování Kvitace pfii pfiekroãení mezní hodnoty Hodnota mezní hodnoty 1 Hodnota mezní hodnoty 2
RST INPT DEC.P RD.S.C RD.CF S1.CF S2.CF S1.DB S2.DB OT.CF + OL - OL RS.OF 9999 - 1999 ER 1 T.RS SP.RS SP 1 SP 2
Klávesnice blokována RESET Input Decimal Point
14.0 Specifikace Vstupy mûfieného signálu +V
Vstupní rozsah Izolaãní napûtí NMR CMR Max. vstupní napûtí Vstupní odpor
-V
-6-
0-100 mV, 650 mV, 0-10 V, 65 V, 0-20 mA, 4-20 mA 354 V vrcholové podle IEC pfiedpisÛ 60 dB 120 dB Max 240 V ef pro napûÈov˘ rozsah signálu; 200 mA pro proudov˘ rozsah vstupu 100 MOhm pro rozsah 10 V nebo 65 V 5 Ohm pro rozsah 20 mA
- 15 -
8.0 Pfiipojení analogového v˘stupu
Menu M.RST Kompenzace odchylky od Ïádané hodnoty pfii P-regulaãní funkci analogového v˘stupu. Tato poloÏka menu se objeví jen pfii nastavení 0.3=P. Jako hodnota bude programovaná odchylka mezi mûfienou hodnotou a mezní hodnotou 1 (SP1). MENU
Volba
Popis
M.RST
0..9999
Odchylka
Pomocí tlaãítek TARE a MAX nastavte hodnotu odchylky mûfiené veliãiny k mezní hodnotû 1 (Ïádaná hodnota). Tlaãítkem TARE mûníte pozice aktivního digitu (blikající ãíslice), pomocí tlaãítka MAX nastavujete ãíselnou hodnotu. Tlaãítkem MENU uloÏíte nastavenou odchylku do pamûti. Stiskem tlaãítka RESET bude vyvolána pfiedchozí poloÏka programovacího menu, pomocí dvojitého stlaãení tlaãítka RESET dojde k ukonãení programování a resetu pfiístroje. napûÈov˘ v˘stup
MENU
Volba
Popis
LK.CF LK.CF
SP=D* SP=E*
zmûna mezních hodnot nepovolena zmûna mezních hodnot povolena
zem
9.0 Pfiipojení reléov˘ch v˘stupÛ relé 1
relé 2
spín 1 rozep 1 COM1 spín 2 rozep 2 COM2
Menu LK.CF, programová blokace pro mezní hodnoty a RESET
proudov˘ v˘stup
relé 1 = mezní hodnota 1 relé 2 = mezní hodnota 2
LK.CF RS=D RESET-tlaãítko nepovoleno v RUN-módu LK.CF RS=E RESET-tlaãítko povoleno v RUN-módu * Prosím pov‰imnûte si v této souvislosti nastavení Hardware-programové blokace (zástrãka S3, Manuál str.4) Stiskem tlaãítka MENU uloÏíte nastavené hodnoty do pamûti pfiístroje. Stiskem tlaãítka RESET bude vyvolána pfiedchozí poloÏka programovacího menu, pomocí dvojitého stlaãení tlaãítka RESET dojde k ukonãení programování a resetu pfiístroje.
10.0 Nastavení mezních hodnot 12. Funkce tlaãítek v RUN-Módu. Pro nastavení mezních hodnot z programového menu musí b˘t propojka S3-A instalována a S3-E odpojena. K nastavení mezní hodnoty musí b˘t nastaven v menu LK.CF parametr SP=E a propojky S3-A a. S3-E instalovány. V této konfiguraci je jednotlivé programové menu blokováno. Pfii stlaãení tlaãítka MENU se objeví na displeji LOCK. Mezní hodnoty se dají stlaãením SETPS vyvolat a pfienastavit. Pokud je parametr v menu LF.CF na SP=D, lze mezní hodnoty stlaãením tlaãítka SEPTS zobrazt na displeji, ale nelze jejich hodnoty mûnit. Nastavení mezní hodnoty: Stisknutím tlaãítka SEPTS se krátce objeví SP 1, poté se na displeji objeví nastavená pfiedchozí nastavená mezní hodnota 1. Pomocí tlaãítek TARE a MAX mÛÏete nastavit novou Ïádanou hodnotu. Tlaãítkem TARE mûníte pozice aktivního digitu (blikající ãíslice), pomocí tlaãítka MAX nastavujete ãíselnou hodnotu.
12.1 Vyvolání uloÏen˘ch MAX-hodnot. Stisknout tlaãítko MAX
Bude se pfiepínat mezi aktuální mûfienou hodnotou a uloÏenou MAXhodnotou. Pro rozli‰ení bude hodnota maxima na displeji blikat.
- 14 -
Tlaãítkem SEPTS uloÏíte novou hodnota do pamûti pfiístroje. Krátce se objeví SP 2, obdobn˘m zpÛsobat mÛÏete i tuto pfienastavit. Pokud bûhem 15-ti sekund nebude stlaãeno Ïádné tlaãítko, vrací se pfiístroj do normálního RUN módu (mûfiení a indikace).
-7-
OT.CF OT.CF Nastavení mûfiícíhorozsahu INPT (INPUT). Po stlaãení tlaãítka MENU se na displeji objeví INPT.Tlaãítkem TARE navolte poÏadovan˘ vstupní rozsah. MENU
Volba
Popis
INPT INPT INPT INPT INPT
100 m 650m 10 V 65 V 0 - 20
Rozsah Rozsah Rozsah Rozsah Rozsah
mûfiení mûfiení mûfiení mûfiení mûfiení
100 mV 650 mV 10 V 65 V 0/4 - 20 mA
Tlaãítkem MAX provedete zmûnu volby vstupního signálu. Volbu potvrìte stisknuím tlaãítka MENU a zvolená hodnota se uloÏí do pamûti pfiístroje. Stiskem tlaãítka RESET bude vyvolána pfiedchozí poloÏka programovacího menu, pomocí dvojitého stlaãení tlaãítka RESET dojde k ukonãení programování a resetu pfiístroje.
0.5=F 0.5=H
P-regulátor analogov˘ v˘stup pracuje s 0-10V P-regulátor analogov˘ v˘stup pracuje s 0-5V
Opakovan˘n stiskem tlaãítka TARE naleznûte pfiíslu‰né menu, pomocí tlaãítka MAX nastavte poÏadované hodnoty. Stisknutím tlaãítka MENU uloÏíte hodnoty do pamûti pfiístroje. Stiskem tlaãítka RESET bude vyvolána pfiedchozí poloÏka programovacího menu, pomocí dvojitého stlaãení tlaãítka RESET dojde k ukonãení programování a resetu pfiístroje.
Menu OT.S.O, pfiifiazení indikace k analogovému v˘stupu Pfiifiazení indikace k analogovému v˘stupu Pfiíklad 1: zobrazení 0-1500, odpovídá analogovému v˘stupu 0-10 V, Pfiíklad 2: zobrazení 0-8000 odpovídá analogovému v˘stupu 4-20 mA. MENU
Volba
Popis
OT.S.0 Pfiíklad 1: 0-1500 = 0-10 V Menu DEC.P, nastavení pozice desetinné teãky MENU
Volba
Popis
DEC.P DEC.P DEC.P DEC.P
FFFF. FFF.F FF.FF F.FFF
Pozice Pozice Pozice Pozice
desetinné desetinné desetinné desetinné
teãky teãky teãky teãky
RD 1 OUT.1 RD 2 OUT.2
0000 00.00 1500 10.00
Indikace Lo=0 Analogov˘ v˘stup Lo=OV Indikace Hi=1500 Analogov˘ v˘stup Hi=10 V
Pfiíklad 2: 0-8000 = 4 - 20 mA
Pozice desetinné teãky nemá v Ïádn˘ vliv na mûfien˘ údaj INFCP, slouÏí pouze ke korektnímu zobrazení mûfiené honoty na displeji ve fyzikální jednotce. Nastavení pozice desetinné teãky se objeví po stlaãení tlaãítka TARE, pomocí tlaãítka MAX mÛÏete desetinnou teãku pfiesunout na poÏadovanou pozici. Pomocí tlaãítka MENU pak zvolenou pozici uloÏíte do pamûti pfiístroje. Stiskem tlaãítka RESET bude vyvolána pfiedchozí poloÏka programovacího menu, pomocí dvojitého stlaãení tlaãítka RESET dojde k ukonãení programování a resetu pfiístroje.
RD 1 OUT.1 RD 2 OUT.2
0000 04.00 8000 20.00*
Indikace Lo=0 Analogov˘ v˘stup Lo=4 mA Indikace Hi=8000 Analog.v˘stup Hi=20 mA
maximální hodnota v˘stupního proudového signálu je 20 mA Nastavení analogového v˘stupu uloÏite do pamûti pomocí tlaãítka MENU. Stiskem tlaãítka RESET bude vyvolána pfiedchozí poloÏka programovacího menu, pomocí dvojitého stlaãení tlaãítka RESET dojde k ukonãení programování a resetu pfiístroje.
Menu P.BND.
11.0 Pro programování INFCP musí b˘t zástrãka S3A instalována a S3-E odpojena
Tato poloÏka menu se objeví jen pfii nastavení 0.3=P /Analogov˘ v˘stup pracuje jako P-regulátor). Tímto menu P.BND nastavíte pásmo proporcionality nalogového v˘stupu. MENU
Volba
Popis
Menu RD.S.O. (Reading Scale and Offset), nastavení zobrazení
P.BAND
0..9999
Pásmo proporcionality analogového v˘stupu
Metoda 1: Aktivní signál z kalibrátoru nebo snímaãe na vstupu INFCP Pfiíklad: Vstupní signál cca 0 - 10 V musí odpovídat indikaci od 0 - 8500.
Pomocí tlaãítek TARE a MAX bude nastavte poÏadovanou hodnotu pásma proporcionality. Tlaãítkem TARE mûníte pozice aktivního digitu (blikající ãíslice), pomocí tlaãítka MAX nastavujete ãíselnou hodnotu.
MENU
Stisknutím tlaãítka MENU uloÏíte pfiedchozí nastavení pásma proporcionality uloÏeno.
Popis
RD.S.
Stiskem tlaãítka RESET bude vyvolána pfiedchozí poloÏka programovacího menu, pomocí dvojitého stlaãení tlaãítka RESET dojde k ukonãení programování a resetu pfiístroje.
-8-
- 13 -
poÏadované hodnoty. Stisknutím tlaãítka MENU uloÏíte hodnoty do pamûti pfiístroje.
1. Pfievzetí mûfiené hodnoty signálu Lo IN 1 (INPUT 1). Vstupní signál Lo musí b˘t pfiipojen
Stiskem tlaãítka RESET bude vyvolána pfiedchozí poloÏka programovacího menu, pomocí dvojitého stlaãení tlaãítka RESET dojde k ukonãení programování a resetu pfiístroje.
Stisknout tlaãítko TARE
Na displeji se objeví IN 1
Stisknout tlaãítko TARE
Objeví se naposledy uloÏená hodnota INPUT 1
Stisknout tlaãítko TARE
Zobrazí se mûfiená hodnota Lo (napfi. 456=0,456 V)
Stisknout tlaãítko MENU
Dojde k uloÏení nové hodnoty do pamûti a na displeji se objeví RD 1
Menu S1.DB, nastavení hysterese mezní hodnoty 1 MENU
Volba
Popis
S1.DB
0..9999
Hysterese pro mezní hodnotu 1
Pomocí tlaãítek TARE a MAX bude nastavte poÏadovanou hodnotu hysterese. Tlaãítkem TARE mûníte pozice aktivního digitu (blikající ãíslice), pomocí tlaãítka MAX nastavujete ãíselnou hodnotu. Stisknutím tlaãítka MENU uoloÏíte hodnotu nastavení hysterese do pamûti. Stiskem tlaãítka RESET bude vyvolána pfiedchozí poloÏka programovacího menu, pomocí dvojitého stlaãení tlaãítka RESET dojde k ukonãení programování a resetu pfiístroje.
2. PoÏadované zobrazení pfii signálu Lo RD 1 (0000) Tlaãítky TARE a MAX nastavte poÏadované zobrazení pfii signálu Lo. Vstup 0000 (indikace pfii 0,456 V = 0). Hodnotu uloÏíte stiskem tlaãítka MENU. 3. Pfievzetí mûfiené hodnoty signálu Hi IN 2 (INPUT 2). Vstupní signál Hi musí b˘t pfiipojen. Stisknout tlaãítko TARE Stisknout tlaãítko TARE Stisknout tlaãítko MENU
Nastavení mimo provoz (Ïádná hysterese)
Objeví se naposledy uloÏená hodnota INPUT 2 Zobrazí se mûfiená hodnota Hi (napfi. 9875 = 9,875 V) Dojde k uloÏení nové hodnoty do pamûti a na displeji se objeví RD 2
4. PoÏadované zobrazení pfii signálu Hi RD 2 (8500) Menu S2.DB, nastavení hysterese mezní hodnoty 2 MENU
Volba
Popis
S2.DB
0..9999
Hysterese pro mezní hodnotu 2
Pomocí tlaãítek TARE a MAX bude nastavte poÏadovanou hodnotu hysterese. Tlaãítkem TARE mûníte pozice aktivního digitu (blikající ãíslice), pomocí tlaãítka MAX nastavujete ãíselnou hodnotu. Stisknutím tlaãítka MENU uoloÏíte hodnotu nastavení hysterese do pamûti. Stiskem tlaãítka RESET bude vyvolána pfiedchozí poloÏka programovacího menu, pomocí dvojitého stlaãení tlaãítka RESET dojde k ukonãení programování a resetu pfiístroje.
Tlaãítky TARE a MAX nastavte poÏadované zobrazení pfii signálu Hi. Vstup 8500 (indikace pfii 9.875 V = 8500). Hodnotu uloÏíte stiskem tlaãítka MENU. V pfiípadû, Ïe pfii ‰kálování pfiipojíte signál napfi. 21 mA, objeví se na displeji blikající 9999. Vstupní hodnoty, které leÏí max. 10% nad specifikovan˘ vstupní rozsah, budou pfiesto pfievzaty. Pfiípadnû zobrazená desetinná teãka nemá Ïádn˘ vliv na ‰kálování. Pfii prÛbûhu ‰kálování musí b˘t popsan˘ postup exaktnû dodrÏen. Pomocí tlaãítka RESET se mÛÏete pfii chybném zadání nûkteré z hodnot vrátit zpût na menu RD.S.O. a provést celou proceduru znovu.
Nastavení mimo provoz (Ïádná hysterese)
Nastavením interního rozli‰ení R2=4 (viz pfií‰tí programov˘ bod RD.DF) obdrÏíte se pfii pfiipojen˘ch vstupních signálech INPUT 1 a INPUT 2 následující hodnoty:
Menu OT.CF, konfigurace analogového v˘stupu
Vstupní signál 0 - 100 mV -50..+50 mV 0 - 10 V
MENU
Volba
Popis
OT.CF OT.CF
0.1=D 0.1=E
Analogov˘ v˘stup nepovolen Analogov˘ v˘stup povolen
OT.CF OT.CF
0.2=V 0.2=C
Analogov˘ v˘stup 0-10 V (napûÈov˘) Analogov˘ v˘stup 0/4 - 20 mA (proudov˘)
OT.CF OT.CF
0.3=A 0.3=P
Analogov˘ v˘stup pracuje v normálním provozu Analogov˘ v˘stup pracuje jako P-regulátor (proporcionální)
Následující menu se objeví jen pfii nastavení 0.3=P OT.CF OT.CF
0.4=D 0.4=R
P-regulátor analogov˘ v˘stup pracuje v pfiímé funkci P-regulátor analogov˘ v˘stup pracuje v reverzní funkci
Následující menu se objeví jen pfii nastavení 0.2= V a 0.3=P
- 12 -
Indikace 0 - 9999 -2000..+2000 0 - 9999
Vstupní signál -5V ..+5V 0 - 20 mA 4 - 20 mA
Indikace -2000..+2000 0 - 9999 2000 - 9999
Metoda 2: V˘stupní data snímaãe jsou známá, pfiesto nemohou b˘t simulována. Pfiíklad: vstupnímu napûtí 0,55 - 21,48 mV musí odpovídat indikace 0 - 6000 MENU RS.S.0 Popis 1. Vstup signálu-Lo mûfiené hodnoty IN 1 (INPUT 1) Stisknout tlaãítkoTARE Stisknout tlaãítkoTARE
Na displeji se objeví IN 1 Objeví se naposledy uloÏená hodnota INPUT 1
Pomocí tlaãítek TARE a MAX nastavte hodnotu signálu-Lo. Vstup 0055 (0,55 mV) pro signál Lo
-9-
Hodnotu uloÏíte stiskem tlaãítka MENU a na displeji se objeví RD 1 2. Zobrazení na displeji pfii signálu-Lo RD 1 (0000) Pomocí tlaãítek TARE a MAX nastavte poÏadované zobrazení odpovídající signálu-Lo. Vstup 0000 (indikace pfii 0,55 mV = 0000) Hodnotu uloÏíte stiskem tlaãítka MENU a na displeji se objeví IN 2 3. Vstup signálu-Hi mûfiené hodnoty IN 2 (INPUT 2)
PoÏadovan˘ parametr zvolíte pomocí tlaãítka TARE, tlaãítkem MAX nastavte poÏadovanou hodnotu. Stisknutím tlaãítka MENU uloÏíte do pamûti zvolen˘ parametr. Stiskem tlaãítka RESET bude vyvolána pfiedchozí poloÏka programovacího menu, pomocí dvojitého stlaãení tlaãítka RESET dojde k ukonãení programování a resetu pfiístroje. Interní rozli‰ení: rozli‰ení pfii Vstupní rozsah R0 R1 R2 R3 R4
Pomocí tlaãítek TARE a MAX nastavte hodnotu signálu-Hi. Vstup 2148 (21,48 mV) pro signál Hi Hodnotu uloÏíte stiskem tlaãítka MENU a na displeji se objeví RD 2
0-100 mV 650 mV 0 - 10 V 65 V 0/4-20mA
4. Zobrazení na displeji pfii signálu-Hi RD 2 (6000)
Pfiíklad 1 (vstupní signál 0 - 1 V)
Pomocí tlaãítek TARE a MAX nastavte poÏadované zobrazení odpovídající signálu-Hi. Vstup 6000 (indikace pfii 21,48 mV = 6000) Hodnotu uloÏíte stiskem tlaãítka MENU
Pfii vstupních signálech v koncovém rozsahu napfi. mûfien˘ rozsah 0 - 10 V musí b˘t obecnû zvoleno nastavení R4. V pfiípadû, Ïe napfi. mûfien˘ signál má jen 1V, mÛÏe pfii rozsahu 10 V b˘t zvoleno nastavení R0. S R0 bude mûfien˘ signál mûfien s desetinásobn˘m rozli‰ením (indikace pfii 1 V = 9999). S vût‰ím rozli‰ením se je moÏno nastavit pfiesnûji ‰kálování.
Stisknout tlaãítkoTARE
Objeví se naposledy uloÏená hodnota INPUT 2
Pomocí tlaãítka RESET se mÛÏete pfii chybném zadání nûkteré z hodnot vrátit zpût na menu RD.S.O. Pfiípadnû zobrazená desetinná teãka nemá Ïádn˘ vliv na ‰kálování. Desetinná teãka slouÏí u pfiístrojové fiady INFCP pouze ke správnému zobrazení ve fyzikálních jednotkách. Hodnoty na displeji pfii IN 1 (INPUT 1) a IN (INPUT 2). Uvedené hodnoty se vztahují na nastavení interního rozli‰ení R 2 = 4 (viz. následující menu RD.CF). Vstupní signál 0 - 100 mV -50..+50 V 0 - 10 V -5V..+5 V 0 - 20 mA 4 - 20 mA
Indikace 0 - 9999 -2000..+2000 0 - 9999 -2000..+2000 0 - 9999 2000 - 9999
Menu RD.CF, aktivace funkce tarování a nastavení interního rozli‰ení. MENU
Volba
Popis
RD.CF RD.CF
R1=T R1=N
funkce tarování aktivní funkce tarování neaktivní
10 mV 50 mV 1V 5V 2 mA
9999 (1mV) 5 mV 100 mV 500 mV 0,2 mA
5000 (2mV) 10 mV 200 mV 1 mV 0,4 mA
3333 (3mV) 15 mV 300 mV 1,5 mV 0,6 mA
2000 (5mV) 25 mV 500 mV 2,5 mV 1 mA
1000 (10mV) 25 mV 1 mV 2,5 mV 2 mA
Menu S1.CF, Funkce mezní hodnota 1 Menu
Volba
Popis
S1.CF
S.1=A
Aktivní nad nastavenou hodnotou SP1
S.1CF
S.1=B
Aktivní pod nastavenou hodnotou SP1
S.1CF S.1CF
S.2=U S.2=L
Mezní hodnota 1 pracuje v nepotvrzovacím provozu Mezní hodnota 1 pracuje v potvrzovacím provozu (pfiekroãení mezní hodnoty musí b˘t potvrzeno operátorem)
Opakovan˘n stiskem tlaãítka TARE naleznûte pfiíslu‰né menu, pomocí tlaãítka MAX nastavte poÏadované hodnoty. Stisknutím tlaãítka MENU uloÏíte hodnoty do pamûti pfiístroje. Stiskem tlaãítka RESET bude vyvolána pfiedchozí poloÏka programovacího menu, pomocí dvojitého stlaãení tlaãítka RESET dojde k ukonãení programování a resetu pfiístroje.
Interní rozli‰ení pfii napûÈov˘ch vstupech unipolárních RD.CF RD.CF RD.CF RD.CF RD.CF RD.CF RD.CF
R2=0 R2=1 R2=2 R2=3 R2=4 R3=F R3=U
bipolárních
1 mV 5 mV 2 mV 10 mV 3 mV 15 mV 5 mV 25 mV 10 mV 25 mV Inteligentní filtr pro indikaci aktivní Inteligentní filtr pro indikaci neaktivmní
- 10 -
Menu S2.CF, funkce mezní hodnota 2 MENU
Volba
Popis
S2.CF S2.CF
S.1=A S.1=B
Aktivní nad nastavenou hodnotou SP2 Aktivní pod nastavenou hodnotou SP2
S2.CF S2.CF
S.2=U S.2=L
Mezní hodnota 2 pracuje v nepotvrzovacím provozu Mezní hodnota 2 pracuje v potvrzovacím provozu (pfiekroãení mezní hodnoty musí b˘t potvrzeno operátorem)
Opakovan˘n stiskem tlaãítka TARE naleznûte pfiíslu‰né menu, pomocí tlaãítka MAX nastavte
- 11 -
INFCP and INFCP-xxxB INFINITY® C Process Panel Meter Operator’s Manual
NEWPORT Electronics, Inc. http://www.newportUS.com/manuals
Additional products from
NEWPORT Electronics, Inc. Rate Meters Timers Totalizers Strain Gauge Meters Voltmeters Multimeters Soldering Iron Testers pH pens pH Controllers pH Electrodes RTDs Thermowells Flow Sensors
Counters Frequency Meters PID Controllers Clock/Timers Printers Process Meters On/Off Controllers Recorders Relative Humidity Transmitters Thermocouples Thermistors Wire
For Immediate Assistance In the U.S.A. and Canada: 1-800-NEWPORT® In Mexico: (95) 800-NEWPORTSM Or call your local NEWPORT Office. Internet e-mail
[email protected]
NEWPORTnet On-Line Service http://www.newportUS.com SM
It is the policy of NEWPORT to comply with all worldwide safety and EMC/EMI regulations that apply. NEWPORT is constantly pursuing certification of its products to the European New Approach Directives. NEWPORT will add the CE mark to every appropriate device upon certification. The information contained in this document is believed to be correct but NEWPORT Electronics, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING: These products are not designed for use in, and should not be used for, patient connected applications. TRADEMARK NOTICE: a , NEWPORT , NEWPORT , newportUS.com™, Case Bezel Design” are trademarks of NEWPORT Electronics, Inc. ®
®
®
,
,
and the “Meter
PATENT NOTICE: This product is covered by one or more of the following patents: U.S. Pat. No. Des. 336,895; 5,274,577; 6,243,021 / Canada 2052599; 2052600 / Italy 1249456; 1250938 / France Brevet No. 91 12756 / Spain 2039150; 2048066 / UK Patent No. GB2 249 837; GB2 248 954 / Germany DE 41 34398 C2. Other US and International Patents Pending. This device is marked with the international caution symbol. It is important to read the Setup Guide before installing or commissioning this device as it contains important information relating to safety and EMC.
Table of Contents PREFACE Manual Objectives This manual shows you how to set up and use the Programmable Digital Meter.
Standard Procedures: * * * * * * * * * * * * * *
Checking voltage jumpers, or changing voltage power Mounting the panel Selecting the input type Selecting a decimal point position Scaling with known loads (on-line calibration) Scaling without known loads Enabling/disabling the front-panel tare Displaying the filtered/unfiltered input signal Selecting a display color Setting the setpoint's active band Selecting a latched or unlatched operation Setting setpoint deadbands Enabling/disabling setpoint changes Enabling/disabling the RESET button in the Run Mode
Optional Procedures: * * * * * * *
Setting input resolution Enabling/disabling analog output Selecting analog output as current or voltage Selecting analog output or proportional control Selecting proportional band Using manual reset (offsetting setpoint errors) Scaling analog output
For first-time users: Refer to the QuickStart Manual for basic operation and setup instructions.
i
Table of Contents Table A-1. Sections of the Manual If you want to read about:
ii
Refer to section
Unpacking; safety considerations
1
Introduction
Meter description and features
2
About the Meter
Main board power jumpers; panel mounting, sensor input, main power and analog and relay output
3
Getting Started
Input type; decimal point position; reading scale & offset; reading configuration; display color; setpoint configuration; setpoint deadbands; output configuration (analog output); proportional band; manual reset; analog output scaling; lock out configuration; display brightness
4
Configuring the Meter
Display messages
5
Display Messages
Meter menu/sub-menu messages
6
Menu Configuration
Setpoint configuration messages
7
Setpoint Configuration Displays
Specifications
8
Specifications
Factory Preset Values
9
Factory Default Setup as Shipped
Table of Contents Table of Contents Section
Page
SEC 1 1.1 1.2
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1 Safety Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
SEC 2 2.1 2.2 2.3 2.4 2.5 2.6
ABOUT THE METER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 Available Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 Front of the Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 Back of the Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8 Disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
SEC 3 3.1 3.2 3.3 3.4 3.5 3.6 3.7
GETTING STARTED . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11 Rating/Product Label . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11 Main Board Power Jumpers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11 Panel Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14 Connecting Sensor Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15 Connecting Main Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18 Conecting External Tare Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20 Connecting Analog and Relay Outputs . . . . . . . . . . . . . . . . . . . . . . . . . .20
SEC 4 4.1 4.2 4.3 4.3.1 4.3.2 4.4 4.4.1 4.4.2 4.4.3
CONFIGURING THE METER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22 Selecting the Input Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22 Selecting a Decimal Point Position . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23 Selecting Reading Scale and Offset . . . . . . . . . . . . . . . . . . . . . . . . . . . .23 Scaling with Known Loads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24 Scaling without Known Loads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27 Using Reading Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28 Enabling or Disabling the Front-Panel Tare . . . . . . . . . . . . . . . . . . . . . . .29 Setting Input Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29 Displaying the Filtered/Unfiltered Input Signal . . . . . . . . . . . . . . . . . . . . .30
iii
Table of Contents Table of Contents Section
Page
4.5 4.6 4.6.1 4.6.2 4.7 4.7.1 4.7.2 4.8 4.9 4.10 4.10.1 4.10.2 4.10.3 4.11 4.12 4.13 4.14 4.14.1 4.14.2 4.14.3
Selecting a Display Color . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30 Using Setpoint 1 Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31 Setting Setpoint 1's Active Band . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31 Selecting if Setpoint 1 is Latched or Unlatched . . . . . . . . . . . . . . . . . . . .31 Using Setpoint 2 Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32 Setting Setpoint 2's Active Band . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32 Selecting if Setpoint 2 is Latched or Unlatched . . . . . . . . . . . . . . . . . . . .32 Setting the Setpoint 1 Deadband . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33 Setting the Setpoint 2 Deadband . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33 Using Output Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35 Enabling or Disabling the Analog Output . . . . . . . . . . . . . . . . . . . . . . . . .35 Selecting Analog Output as Current or Voltage . . . . . . . . . . . . . . . . . . . .36 Selecting Analog Output or Proportional Control . . . . . . . . . . . . . . . . . . .36 Selecting Proportional Band . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38 Using Manual Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40 Using Output Scale and Offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41 Using Lock Out Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43 Enabling or Disabling the RESET Button in the Run Mode . . . . . . . . . . .43 Enabling or Disabling Setpoint Changes . . . . . . . . . . . . . . . . . . . . . . . . .43 Setpoints Display Function: Software Version or Setpoint Value . . . . . . . . . . . . . . . . . . . . . . . . . . . .44 4.15 Using Display Brightness Configuration . . . . . . . . . . . . . . . . . . . . . . . . .44 4.15.1 Changing Brightness Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .44 SEC 5
DISPLAY MESSAGES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .45
SEC 6
MENU CONFIGURATION DISPLAYS . . . . . . . . . . . . . . . . . . . . . . . . . .46
SEC 7
SETPOINT CONFIGURATION DISPLAYS . . . . . . . . . . . . . . . . . . . . . . .51
SEC 8
SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .52
SEC 9
FACTORY PRESET VALUES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57 CE APPROVAL INFORMATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .58
iv
Table of Contents List of Figures Figure 2-1 2-2 2-3 3-1 3-2 3-3 3-4 3-5 3-6 3-7 3-8 3-9 3-10 3-11 3-12 3-13 3-14 3-15 3-16 3-17 3-18 4-1 4-2 8-1
Page Front-Panel with “Big” LED Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 Front-Panel with Standard LED Display . . . . . . . . . . . . . . . . . . . . . . . . . .5 Connector Label (AC-Powered and DC-Powered Detail) . . . . . . . . . . . . .8 Main Board Power Jumpers (W1, W2, W3) . . . . . . . . . . . . . . . . . . . . . .11 Main Board Jumper Positions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12 Upper Isolated Analog Output Option Board Installation . . . . . . . . . . . . .12 Meter - Exploded View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14 Panel Cut-Out . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14 3-Wire DC Input Connections with Internal Excitation . . . . . . . . . . . . . . .15 3-Wire DC Input Connections with External Excitation . . . . . . . . . . . . . .15 4-Wire DC Input Connections with Internal Excitation . . . . . . . . . . . . . . .16 4-Wire DC Input Connections with External Excitation . . . . . . . . . . . . . .16 DC Current Input Connectiions with Internal Excitation . . . . . . . . . . . . . .17 DC Current Input Connections with External Excitation . . . . . . . . . . . . .17 DC Current Input Connections with Current Source . . . . . . . . . . . . . . . .18 Main Power Connections - AC Powered Unit . . . . . . . . . . . . . . . . . . . . .18 Main Power Connections - DC Powered Unit . . . . . . . . . . . . . . . . . . . . .19 External Tare Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20 Analog Output Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20 Relay Output Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21 Isolated Analog Output Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 Alarm Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .34 Controller Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38 Meter Dimensions/Panel Cutout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .56
v
Table of Contents List of Tables Table A-1 2-1 2-2 2-3 3-1 3-2 4-1 4-2 4-3 4-4 5-1 6-1 6-2 7-1 8-1 9-1
vi
Page Sections of the Manuals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ii Accessories and Add-Ons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 Connector Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9 DIP Switch Positions/Input Range & Excitation . . . . . . . . . . . . . . . . . . . .10 S3 Jumper Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13 Main Power Connection - AC-Powered Unit . . . . . . . . . . . . . . . . . . . . . .19 Range Selection Dip Switch Positions for Regular Voltage Input . . . . . .24 Range Selection Dip Switch Positions for Millivolt and Milliamp Input . . .24 Natural Gain . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27 Input Resolution Multiplier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27 Display Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .45 Menu Configuration Displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .46 Run Mode Displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50 Setpoint Configuration Displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .51 Color Chart for dc Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .55 Factory Preset Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57
Notes, Warnings and Cautions NOTES, WARNINGS and CAUTIONS Information that is especially important to note is identified by three labels: • • • • Note ☞
NOTE WARNING CAUTION IMPORTANT
NOTE: provides you with information that is important to successfully setup and use the Programmable Digital Meter. CAUTION or WARNING: tells you about the risk of electric shock.
CAUTION, WARNING or IMPORTANT: tells you of circumstances or practices that can effect the meter's functionality and must refer to accompanying documents. TIP: Provides you helpful hints.
vii
viii
Introduction
1
SECTION 1. INTRODUCTION 1.1 UNPACKING Remove the Packing List and verify that all equipment has been received. If there are any questions about the shipment, use the phone numbers listed on the back cover to contact the Customer Service Department nearest you. Upon receipt of shipment, inspect the container and equipment for any signs of damage. Take particular note of any evidence of rough handling in transit. Immediately report any damage to the shipping agent.
Note ☞
The carrier will not honor any claims unless all shipping material is saved for their examination. After examining and removing contents, save packing material and carton in the event reshipment is necessary.
Verify that you receive the following items in the shipping box: QTY
DESCRIPTION
1
Programmable Digital Meter indicator/controller with all applicable connectors attached.
1
Owner's Manual
1
Set Mounting brackets
Note ☞
If you ordered any of the available options (except the "BL" Blank Lens option), they will be shipped in a separate container to avoid any damage to your indicator/controller.
1
1
Introduction 1.2
SAFETY CONSIDERATIONS This device is marked with the international caution symbol. It is important to read this manual before installing or commissioning this device as it contains important information relating to Safety and EMC (Electromagnetic Compatibility). This instrument is a panel mount device protected in accordance with EN 610101:2001, electrical safety requirements for electrical equipment for measurement, control and laboratory. Installation of this instrument should be done by qualified personnel. In order to ensure safe operation, the following instructions should be followed. This instrument has no power-on switch. An external switch or circuit-breaker shall be included in the building installation as a disconnecting device. It shall be marked to indicate this function, and it shall be in close proximity to the equipment within easy reach of the operator. The switch or circuit-breaker shall not interrupt the Protective Conductor (Earth wire), and it shall meet the relevant requirements of IEC 947–1 and IEC 947-3 (International Electrotechnical Commission). The switch shall not be incorporated in the main supply cord. Furthermore, to provide protection against excessive energy being drawn from the main supply in case of a fault in the equipment, an overcurrent protection device shall be installed. •
Note ☞
• • • • • •
Do not exceed voltage rating on the label located on the top of the instrument housing. Always disconnect power before changing signal and power connections. Do not use this instrument on a work bench without its case for safety reasons. Do not operate this instrument in flammable or explosive atmospheres. Do not expose this instrument to rain or moisture. Unit mounting should allow for adequate ventilation to ensure instrument does not exceed operating temperature rating. Use electrical wires with adequate size to handle mechanical strain and power requirements. Install without exposing bare wire outside the connector to minimize electrical shock hazards.
EMC Considerations • Whenever EMC is an issue, always use shielded cables. • Never run signal and power wires in the same conduit. • Use signal wire connections with twisted-pair cables. • Install Ferrite Bead(s) on signal wires close to the instrument if EMC problems persist. Failure to follow all instructions and warnings may result in injury! 2
About The Meter
2
SECTION 2. ABOUT THE METER 2.1 DESCRIPTION The Digital Programmable Process meter is a value packed indicator/controller. Four full digits and broad scaling capability allow for display in virtually all engineering units. A wide variety of DC current and voltage input ranges cover typical process applications. Standard features include sensor excitation and front panel or remote tare. Your meter may be a basic indicator or it may include analog output or dual relay output. Analog or dual relay output must be ordered at time of purchase. Analog output is fully scalable and may be configured as a proportional controller, or to follow your display. Dual 5 amp, form C relays control critical processes. A mechanical lockout has been included to guard against unauthorized changes.
2.2 FEATURES The following is a list of standard features: * * * * * * * * *
4-digit three color Programmable “Big” LED display or 4-digit, Standard LED Display NEMA 4 / Type 4 Front Bezel ±0.03 % accuracy 8 DC input ranges: 0-100 mV, ±50 mV, 0-5 V, 1-5 V, 0-10 V, ±5 V, 0-20 mA, and 4-20 mA 5, 10, 12, or 24 Vdc sensor excitation Peak detection Front panel and remote tare function Nonvolatile memory-no battery backup 115 or 230 Vac 50/60 Hz power supply or 10-32 Vdc or 26-56 Vdc
The following is a list of optional features: * * * *
Note ☞
Dual 5 amp, form C relay outputs Scalable analog output Proportional control Easy setup for proportional control Features with are for the “B” version which has three-color programmable “Big” LED display - All segment characters shown are for the “B” version. 3
2
About The Meter 2.3 AVAILABLE ACCESSORIES
Table 2-1. Accessories and Add-Ons Add-On Options FS SPC4 SPC18
Special Calibration/Configuration NEMA-4 Splash Proof Cover NEMA-4 Splash Proof Cover, NEW
Accessories TP1A RP18 RP28 RP38
4
Trimplate panel adaptor. Adapts DIN1A/DIN2A cases to larger panel cutouts 19-In. Rack Panel for one (1) 1/8 DIN instrument 19-In. Rack Panel for two (2) 1/8 DIN instruments 19-In. Rack Panel for three (3) 1/8 DIN instruments
About The Meter
2
2.4 FRONT OF THE METER Figure 2-1 shows each part of the front of the three-color programmable “Big” LED display meter (Version B).
Digital LED Display: -1.9.9.9 or 9.9.9.9 4-digit three color programmable, 21 mm (0.83") high LED display with programmable decimal point.
Figure 2-1. Front-Panel with Big Display
Figure 2-2 shows each part of the front of the standard LED display meter.
•••• 1
SETPTS
Digital LED Display: -1.9.9.9. or 9.9.9.9. 14 segment, 13.8 mm (0.54") high LED display with programmable decimal point.
2
/MAX
/TARE
MENU
RESET
Figure 2-2. Front-Panel wtih Standard Display These meter display windows (both versions) light when appropriate: 1 - Setpoint 1 status 2 - Setpoint 2 status 5 Pushbuttons for programming the meter. 5
2
About The Meter 2.4 FRONT OF THE METER(Continued) METER BUTTONS SETPTS Button In the Run Mode, this button will sequentially recall the previous setpoint settings. As necessary, use the /MAX and /TARE buttons to alter these settings, then press the SETPTS button to store new values. Unless you press the SETPTS, /TARE, or /MAX button within 20 seconds, the meter will scroll to setpoint 2 and then to the Run Mode.
Note ☞
If the dual relay option is not installed or if the L.3=1 on the LK.CF menu, pressing the SETPTS button will display the meter's firmware version.
/MAX Button
In the Run Mode, this button will recall the PEAK reading since the last press of the RESET button. In the Configuration Mode, press this button to change the value of the flashing digit shown on the display and/or toggle between menu choices, such as R.1=T or R.1=N on RD.CF menu. When configuring your setpoint values, press the /MAX button to advance the flashing digit's value from 0 to 9 by 1.
/TARE Button
In the Run Mode press the /TARE button to tare your reading (zeroing) if you configure the Reading Configuration bit R.1=T of the RD.CF menu. If you configure R.1=N , the /TARE button has no function. In the Configuration Mode, press the this button to scroll to the next digit.
6
About The Meter
2
2.4 FRONT OF THE METER(Continued) MENU Button In the Run Mode, press the MENU button to terminate the current measuring process and enter you into the Configuration Mode. Note ☞
Only if you have not installed the lockout jumpers on the main board.
In the Configuration Mode, press the MENU button to store changes in the nonvolatile memory and then advance you to the next menu item.
RESET Button If you hard reset (press the MENU button followed by the RESET button) or power off/on the meter, it shows RST , followed by PROC . In the Run Mode, press the RESET button to reset tare, if any. The meter shows T.RST and returns to the Run Mode. In the Configuration Mode, press the RESET button once to review the previous menu. Press the RESET button twice to perform a hard reset and return to the Run Mode. In the Peak Mode, press the RESET button to reset peak values. The meter shows PK.RS and returns to the Run Mode. In the Setpoint Mode, press the RESET button to reset the latched setpoint. The meter shows SP.RS and enters the Run Mode.
Note ☞
When in setpoint or Configuration Mode, if the meter shows 9999 or -1999 with all flashing digits, the value has overflowed. Press the /MAX button to start a new value.
7
2
About The Meter 2.5 BACK OF THE METER Figure 2-2 shows the label describing the connectors on the back of the meter. Table 2-2 on the following page gives a brief description of each connector at the back of the meter.
Figure 2-3. Connector Label (AC-Powered and DC-Powered Detail)
8
About The Meter
2
2.5 BACK OF THE METER (Continued) Table 2-2. Connector Description Connector TB1-1 TB1-2 TB1-3 TB1-4 TB1-5 TB1-6 TB1-7 TB1-8 TB1-9 TB1-10 TB1-11 TB1-12 TB2-1 TB2-2 TB2-3 TB2-4 TB2-5 TB2-6 TB2-7 TB2-8 TB5-1 TB5-2 TB5-3 J1 (1-2)
Description Setpoint 1: Normally open (N.O.1) connection Setpoint 1: Normally closed (N.C.1) connection Setpoint 1: Common (COM1) connection Setpoint 2: Normally open (N.O.2) connection Setpoint 2: Normally closed (N.C.2) connection Setpoint 2: Common (COM2) connection AC line connection (no connections on DC-powered units) AC neutral connection (+ Input on DC-powered units) AC earth ground (DC-power return on DC-powered units) Analog voltage output Analog current output Analog return -E: Negative excitation connection from meter (5, 10, 12 V) +E: Positive excitation connection from meter (5, 10, 12 V) +20 mA connection for analog input Not used. +24 V output connection +S: Positive signal input -S: Negative signal input and return for +20 mA or +24 V Not used Isolated Analog Voltage Output Isolated Analog Current Output Isolated Analog Output Return Remote tare connection with a momentary switch
9
2
About The Meter The DIP switches are located at the S1 position (refer to Figure 3-2). Use a small instrument, such as a paper clip, to change the switches from open to closed. Table 2-3 lists DIP switch settings at the S1 position required to complete the setup of your meter.
Table 2-3. DIP Switch Positions/Input Range & Excitation Function C= Closed O= Open
S1 DIP Switch Positions 1 2 3 4
5
6
7
8
Settings for Excitation Voltage Internal 5/10/12 excitation External 5/10/12 excitation Internal 12 Vdc excitation Internal 10 Vdc excitation Internal 5 Vdc excitation 0-100 mV DC ±50 mV DC ±5 Vdc 0-10 Vdc 0-20 mA DC Note ☞
C
-
-
-
-
-
-
-
O
-
-
-
-
O
O
-
C
-
-
-
-
O
O
-
C
-
-
-
-
C
O
-
-
C
C
-
O C C O O
-
-
O O C C O
C Settings for Input Ranges O C O O C O C O O C O O O C C
The display must also be configured to the selected input type after setting the DIP switches (see Section 4.1, Selecting the Input Type)
2.6 DISASSEMBLY You may need to open up the meter for one of the following reasons: • • Note ☞
To check or change the 115 or 230 Vac power jumpers. To install or remove jumpers on the main board.
Disconnect the power supply before proceeding.
To remove and access the main board, follow these steps:
10
•
Disconnect the main power from the meter.
•
Remove the back case cover.
•
Lift the back of the main board upwards and let it slide out of the case.
Getting Started
3
SECTION 3. GETTING STARTED Caution: The meter has no power-on switch, so it will be in operation as soon you apply power. If you power off/on the meter, or perform a hard reset (press the RESET button twice), the meter shows RST , followed by PROC .
3.1 RATING/PRODUCT LABEL This label is located on top of the meter housing (refer to Figure 3-4).
3.2 MAIN BOARD POWER JUMPERS (refer to Figure 3-1) Note ☞
Important: If you want to change the Factory preset jumpers, do the following steps; otherwise go to section 3.3. Warning: Disconnect the power from the unit before proceeding. This device must only be reconfigured by a specially trained electrician with corresponding qualifications. Failure to follow all instructions and warnings may result in injury!
1.
Remove the main board from the case. Refer to Section 2.6.
2.
Locate the solder jumpers W1, W2, and W3 (located near the edge of the main board alongside the transformer).
3.
If your power requirement is 115 Vac, solder jumpers W1 and W3 should be wired, but jumper W2 should not. If your power requirement is 230 Vac, solder jumper W2 should be wired, but jumpers W1 and W3 should not. Note: W4 jumper is not used.
Figure 3-1 shows the location of solder jumpers W1 through W3.
T1A
W2
W1
DISPLAY BOARD
W3
Figure 3.1 Main Board Power Jumpers 11
3
Getting Started 3.2 MAIN BOARD POWER JUMPERS (Continued) Figure 3-2 shows the location jumper positions on the main board.
DIS PL AY
BO AR D
T1 S3
MAIN BOARD TP1 TP2 TP3 TP4 TP5 TP6 TP7 TP8 TP9 TP10
S4
{
S2
{
TB1
TB2
S1
J1
A B
E D CB A
S3
Refer to table 3-1 S2 C D
TP11
Figure 3-2. Main Board Jumper Positions
Attach cable to P1. TB5
P1
Figure 3-3. Upper Isolated Analog Output Option Board Installation 12
Getting Started
3
3.2 MAIN BOARD POWER JUMPERS (Continued) S2 jumpers are used for testing purposes. Do not use as reading errors may result. S3 jumpers are used for the following (refer to Figure 3-2): * * * *
To enable or disable the front panel push-buttons To allow for an extremely low resistance load for analog output To disable the MENU button To perform calibration procedure
Test pins TP1 - TP11 are for testing purposes. Do not use as reading errors may result. S4-A Factory default jumper installed.
Table 3-1. S3 Jumper Functions Jumper S3-A S3-B S3-C S3-D S3-E
Description Install to enable front panel push-buttons. Remove to disable all front panel push-buttons. Removed. Install for factory calibration only. Removed. Not used. Removed. Not used. If installed without S3-B, the MENU button locks out. If you press the MENU button, the meter shows LOCK .
13
3
Getting Started 3.3 PANEL MOUNTING
RE A (R R C EM OV OV ER ED CONNECTOR )
PANEL CUT-OUT "NEW" STYLE MOUNTING BRACKET
LABEL
PRODUCT LABEL
CASE
GASKET
"OLDER" STYLE MOUNTING BRACKET 2 PCS
FRONT BEZEL
Figure 3-4. Meter - Exploded View 1. Cut a hole in your panel, as shown in Figure 3-4. For specific dimensions refer to Figure 3-5. 2. Insert the meter into the hole. Be sure the front bezel gasket is flush to the panel. 3. Slide on mounting bracket to secure. 4. Proceed to Section 3.4 to connect your sensor input and main power.
PANEL THICKNESS 6,4 (.25) MAX 0,8 (.03) MIN
1,5 R(.06) 4 PLCS
45,00 + 0,61/-0,00 (1.772 + .024/–.000)
92,00 + 0,81/–0,00 (3.622 + .032/–.000) NOTE: Dimensions in Millimeters (Inches)
Figure 3-5. Panel Cut-Out
14
Getting Started
3
3.4 CONNECTING SENSOR INPUTS Figures 3-6 through 3-12 describe how to connect your sensors.
TB2
-E
+E +20mA N/C
1
2
3
+24V +S 5
4
-S
N/C
7
8
6
EXC
SIG COM
Figure 3-6. 3-Wire DC Input Connections with Internal Excitation
TB2
-E
+E +20mA N/C
1
2
3
+24V +S 5
EXTERNAL SUPPLY
+ + -
-
6
4
-S
N/C
7
8
+ -
Figure 3-7. 3-Wire DC Input Connections with External Excitation
15
3
Getting Started 3.4 CONNECTING SENSOR INPUTS (Continued)
TB2
-E
+E +20mA N/C
1
2
3
+24V +S 5
6
+
+
4
-S
N/C
7
8
VOLTAGE
-
-
EXCITATION
OUTPUT
Figure 3-8. 4-Wire DC Input Connections with Internal Excitation
TB2
-E
+E +20mA N/C
1
2
3
+24V +S 5 + + EXTERNAL SUPPLY
-
-
6
+ -
4
-S
N/C
7
8
SENSOR OUTPUT
Figure 3-9. Wire DC Input Connections with External Excitation
16
Getting Started
3
3.4 CONNECTING SENSOR INPUTS (Continued)
TB2 -E
+E +20mA N/C
1
2
3
+24V +S 5 4-20mA Transmitter
6
4
-S
N/C
7
8
(+) (-)
Figure 3-10. DC Current Input Connections with Internal Excitation
TB2 -E
+E +20mA N/C
1
2
3
+24V +S 5 4-20mA Transmitter
6
4
-S
N/C
7
8
(-)
- EXTERNAL
(+)
+ SUPPLY
Figure 3-11. DC Current Input Connections with External Excitation
17
3
Getting Started 3.4 CONNECTING SENSOR INPUTS (Continued)
TB2 -E
+E +20mA N/C
1
2
3
+24V +S 5
4
-S
N/C
7
8
6
(+) MAX. 20mA
(-) Figure 3-12. DC Current Input Connections with Current Source 3.5 CONNECTING MAIN POWER Connect the AC main power connections as shown in Figure 3-13.
WARNING: Do not connect AC power to your device until you have completed all input and output connections. This device must only be installed by a specially trained electrician with corresponding qualifications. Failure to follow all instructions and warnings may result in injury!
TB1
TB2
1 2 3 4 5 6 1 2 3 4
7 8 9 10 11 12 5 6 7 8
LINE NEUTRAL EARTH
SWITCH
FUSE (L) LINE (N) NEUTRAL
GREEN WIRE
AC Power EARTH GROUND
Check for proper Earth grounding in the power distribution system (single phase).
Figure 3-13. Main Power Connections - AC Powered Unit 18
Getting Started
3
3.5 CONNECTING MAIN POWER (Continued) Table 3-2 shows the wire color and respective terminal connections for both USA and Europe.
Table 3-2. Main Power Connection - AC Powered Unit TB1 7 8 9
AC POWER
WIRE COLORS EUROPE USA
~ AC Line ~ AC Neutral ~ AC Earth
Brown Blue Green/Yellow
Black White Green
Connect the DC main power connections as shown in Figure 3-14.
TB1
TB2
1 2 3 4 5 6 1 2 3 4
7 8 9 10 11 12 5 6 7 8
+ DC
- DC
DC POWER
When using DC power, refer to the Table 8-1 Color Chart in the Specifications Section for Display Color, Intensity, Excitation Voltage and Current, and Analog Output Isolated Option. Failure to use proper ratings may result in damaging the unit.
Figure 3-14. Main Power Connections - DC Powered Unit
19
3
Getting Started 3.6 CONNECTING EXTERNAL TARE SWITCH Connect external tare connections as shown in Figure 3-15.
TB1
TB2
1 2 3 4 5 6 1 2 3 4 S1 OPEN
7 8 9 10 11 12 5 6 7 8
J1 1 2
CLOSED 3 4
EXTERNAL TARE SWITCH
Figure 3-15. External Tare Connections
3.7 CONNECTING ANALOG AND RELAY OUTPUTS If you have purchased a meter with analog or dual relay or isolated analog output, refer to the following drawings for output connections. TB1
TB2
1 2 3 4 5 6 1 2 3 4
7 8 9 10 11 12 5 6 7 8
+ ANALOG VOLTAGE 0-10V dc
-
+ -
ANALOG CURRENT 0-20mA or 4-20mA
ANALOG RTN
Figure 3-16. Analog Output Connections 20
Getting Started
3
3.7 CONNECTING ANALOG AND RELAY OUTPUTS (Continued) Relay 2 COM2
N.C.2
N.O.2
COM1
N.C.1
N.O.1
} }
Relay 1
NO2
EXTERNAL LOAD
COM2
TB1
N L
FUSE
TB2
1 2 3 4 5 6 1 2 3 4
7 8 9 10 11 12 5 6 7 8
Figure 3-17. Relay Output Connections.
0—10VDC
0—20mA OR 4—20mA
1 2 3
TB1
TB2
TB5
1 2 3 4 5 6 1 2 3 4 7 8 9 10 11 12 5 6 7 8
Figure 3-18. Isolated Analog Output Connections.
21
4
Configuring The Meter SECTION 4. CONFIGURING THE METER Note ☞
Refer to Table 6-1 for a summary list of menu configuration.
4.1 SELECTING THE INPUT TYPE INPT To select your appropriate input type signal, follow these steps: Note ☞
22
Before proceeding, set the input DIP switch settings at the back of your meter. (Refer to Table 2-3).
1.
Press the MENU button. The meter shows INPT .
2.
Press the ∂TARE button. The meter flashes one of the following: •
0-20 (for 4-20 mA dc) (Default)
•
100M (for 0-100 mV dc)
•
±50M (for ±50 mV dc)
•
10V (for 0-10 Vdc)
•
±5V (for ±5 Vdc)
3.
Press the ßMAX button to scroll through available choices.
4.
Press the MENU button to store your choice. The meter momentarily shows STRD , followed by DEC.P (Decimal Point).
Configuring The Meter
4
4.2 SELECTING A DECIMAL POINT POSITION DEC.P Note ☞
Refer to Table 6-1 for a summary list of menu configuration. To select a decimal point display position, follow these steps:
1.
Press the MENU button until the meter shows DEC.P .
2.
Press the ∂TARE button. The meter shows one of the following: •
FFF.F
•
FF.FF
•
F.FFF
•
FFFF (Default)
3.
Press the ßMAX button to scroll between available choices.
4.
Press the MENU button to store your choice. The meter momentarily shows STRD , followed by the next menu RD.S.O (Reading Scale and Offset). Or you can press the RESET button to abort and go back to the DEC.P menu.
4.3 SELECTING READING SCALE AND OFFSET RD.S.0 Note ☞
Refer to Table 6-1 for a summary list of menu configuration.
To scale the meter to show readings in engineering units. There are two methods. One method is to scale with known inputs. Another method is to scale without known inputs: you calculate input values based on the transducer specifications and manually enter them through the keyboard.
23
4
Configuring The Meter 4.3.1 Scaling with Known Loads (On-Line Calibration) For maximum resolution, find the maximum signal that will be applied to the meter input.
Note ☞
•
For regular voltage input, refer to the main body of Table 4-1.
•
For millivolt or milliamp input, refer to the main body of Table 4-2.
Set the DIP switch positions as indicated at the top of either Table 4-1 or 4-2. The numbers 1 through 8 in the top row of either table represent dip switches 1 through 8, and the O, C or X directly below the number indicates the correct position of each switch. •
‘O' Switch should be open or up.
•
‘C' Switch should be closed or down.
•
‘X' Switch is used to control excitation (refer to Table 2-3 to determine correct position of these switches).
Once Dip switches have been positioned correctly, apply power. Proceed to the RD.CF (Reading Configuration) and set R2 equal to the value in the right hand column of the chart.
Table 4-1. Range Selection Dip Switch Positions For Regular Voltage Input 12345678 XCOOOXXC 0 - 10 V 0-5V 0-3V 0-2V 0-1V
12345678 XCOOCXXC ±5 V ±5 V ±3 V ±2 V ±1 V
RD.CF* R2= 4 3 2 1 0
Table 4-2. Range Selection Dip Switch Positions For Millivolt/ Milliamp Input 12345678 12345678 12345678 RD.CF* Table 4-1. Range Selection DipXOCOCXX0 Switch Positions For Regular Voltage Input XOCOOXX0 XOCCOXX0 R2= 0 - 100 mV ±50 mV 0 - 20 mA 4 0 - 50 mV ±50 mV 0 - 10 mA 3 0 - 30 mV ±30 mV 0 - 6 mA 2 Table 4-2. Range Selection Dip Switch Positions 0 - 20 mV ±20 mV 0 - 4 mA 1 For Millivolt and Milliamp Input 0 10 mV ±10 mV 0 2 mA 0 p * Reading Configuration 24
Configuring The Meter
4
4.3.1 Scaling with Known Loads (On-Line Calibration) (Continued) To scale with known inputs: apply known loads to a transducer connected to a meter, or simulate the transducer output with a voltage or current simulator. To scale with known inputs, follow these steps: 1.
Apply a known load equal to approximately 0% of the transducer range.
2.
Press the MENU button until the meter shows RD.S.O .
3.
Press the ∂TARE button. The meter shows IN1 (Input 1). Note ☞
IN1 (Input 1) is the unscaled display reading at minimum input.
4.
Press the ∂TARE button again. The meter shows last stored value for Input 1.
5.
Press the ∂TARE button once more. The meter shows the actual signal being received.
6.
Press the MENU button to store this value as IN1 (Input 1). The meter shows RD1 (Read 1). Note ☞
RD1 (Read 1) is the desired display reading at Input 1.
7.
Press the ∂TARE button. The meter shows the last stored value for Read 1.
8.
Press the ßMAX button to change the value of your digits.
9.
Press the ∂TARE button to scroll horizontally to the next digit.
10.
Press the MENU button to store value as RD1 . The meter shows IN!2 (Input 2). Note ☞
IN!2 (Input 2) is the unscaled display reading at maximum input.
25
4
Configuring The Meter 4.3.1 Scaling with Known Loads (On-Line Calibration) (Continued) 11.
Apply a known load equal to approximately 100% of the transducer range.
12.
Press the ∂TARE button again. The meter shows the last stored value for Input 2.
13.
Press the ∂TARE button once more. The meter shows the actual signal being received.
14.
Press the MENU button to store Input 2 value. The meter shows RD!2 (Read 2). Note ☞
26
RD!2 (Read 2) is the desired display reading at input 2.
15.
Press the ∂TARE button. The meter shows the last stored value for Read 2.
16.
Press the ßMAX button to change the value of your digits.
17.
Press the ∂TARE button to scroll horizontally to the next digit.
18.
Press the MENU button to store value as RD!2 (Read 2). The meter momentarily shows STRD , followed by RD.CF . Meter scaling is now complete.
Configuring The Meter
4
4.3.2 Scaling Without Known Loads To scale without known inputs, calculate input values based on the transducer specifications and manually enter them on the front-panel pushbuttons. The following example assumes a pressure transducer with these specifications: Pressure Range:
0 to 2000 PSI
Output Span: 1 to 5 Vdc 1.
Determine the correct values for IN1 and IN!2 based on the transducer specifications. In most cases, RD1 & RD!2 are equal to the minimum and maximum of the transducer output span. The example assumes RD1 & RD!2 are equal to the pressure range of the transducer (RD1 = 0000 and RD!2 = 2000). Calculate IN1 and IN!2 using the transducer output span and the following equation: IN = (Sensor Output) x (Natural Gain) x (Multiplier).
Table 4-3. Natural Gain Input Range 0 to 100 mV ±50 mV 0 to 10 V ±5V 0 to 20 mA
2.
Span Units Millivolts Millivolts Volts Volts Milliamps
Natural Gain 100 cts/mV 40 cts/mV 1000 cts/V 400 cts/V 500 cts/mA
Determine the multiplier by the Input Resolution setting (R.2 in the RD.CF menu) and the input range selected. Typically R.2=4 is suitable for most applications.
Table 4-4. Input Resolution Multiplier Input Range 0 to 100 mV 0 to 10 V 0 to 20 mA ± 50 mV ±5V
R.2=4 1.000 1.000 1.000 1.000 1.000
R.2=3 2.000 2.000 2.000 1.000 1.000
R.2=2 3.333 3.333 3.333 1.667 1.667
R.2=1 5.000 5.000 5.000 2.500 2.500
R.2=0 10.00 10.00 10.00 5.000 5.000 27
4
Configuring The Meter 4.3.2 Scaling Without Known Loads (Continued) 3.
Determine IN1 & IN!2 input range and resolution. The example selects the 0 to 10 V range and 10 uV resolution (R.2=4 ). Example: IN1 = (1 Volt) x (1000 cts/v) x (1.000) = 1000 IN!2 = (5 Volt) x (1000 cts/v) x (1.000) = 5000 RD1 = 0000 RD!2 = 2000
4.
Press MENU button until the meter shows RD.S.O .
5.
Press the ∂TARE button. The meter shows IN1 .
6.
Press the ∂TARE button again, the meter shows the last Input 1 value, with the fourth digit flashing.
7.
Press the ßMAX button to change the value of your digits.
8.
Press the ∂TARE button to scroll horizontally to the next digit.
9.
Press the MENU button to store this value. The meter shows RD1 .
10.
Press the ∂TARE button. The meter shows the last value for read 1.
Repeat steps 7, 8 and 9 until RD1 , IN!2 and RD!2 have been displayed, verified, changed (if necessary) and stored.
4.4 USING READING CONFIGURATION RD.CF Note ☞
Refer to Table 6-1 for a summary list of menu configuration.
You may use Reading Configuration RD.CF to configure your meter for the following:
28
•
To enable or disable the front panel tare
•
To set the input resolution of your meter
•
To display the filtered/unfiltered signal input value
Configuring The Meter
4
4.4.1 Enabling or Disabling the Front-Panel Tare To enable or disable the front-panel tare, follow these steps: 1.
Press the MENU button until RD.CF displays.
2.
Press the ∂TARE button. The meter shows one of the following: •
R.1=T (Tare enabled) (Default)
•
R.1=N (Tare disabled)
3.
Press the ßMAX button to view last stored selection. Press the ßMAX button to toggle between selections.
4.
Press the ∂TARE button to select input resolution or press the MENU button to store your selections. STRD momentarily displays, followed by COLR menu.
4.4.2 Setting Input Resolution To set the input resolution of your meter, follow these steps: 1.
Press the MENU button until RD.CF displays, then press the ∂TARE button twice. or Press the ∂TARE button from R.1 . One of the following displays (default is R.2=4 ): R.2=4 = R.2=0 = R.2=1 = R.2=2 = R.2=3 =
10 µV for Unipolar inputs. 25 µV for Bipolar inputs 1 µV for Unipolar inputs. 5 µV for Bipolar inputs. 2 µV for Unipolar inputs. 10 µV for Bipolar inputs 3 µV for Unipolar inputs. 15 µV for Bipolar inputs. 5 µV for Unipolar inputs. 25 µV for Bipolar inputs
Example: 3 µV resolution means that if you input 0-30 mV, at 30 mV the display shows 9999 . 2.
Press the ßMAX button to scroll through available selections.
3.
Press the ∂TARE button to display the filtered/unfiltered signal input or press the MENU button to store your selections. STRD momentarily displays, followed by COLR menu.
29
4
Configuring The Meter 4.4.3 Displaying the Filtered/Unfiltered Input Signal To display the filtered/unfiltered signal input, follow these steps: 1.
Press the MENU button until RD.CF displays, then press the ∂TARE button three times. or Press the ∂TARE button from R.2. One of the following displays: •
R.3=F (Filtered value) (Default)
•
R.3=U (Unfiltered value)
2.
Press the ßMAX button to toggle between available choices.
3.
Press the MENU button to store your selections. STRD momentarily displays, followed by COLR menu.
4.5 Selecting a Display Color COLR Note ☞
Refer to Table 6-1 for a summary list of menu configuration.
Selecting “Display Color” is not active unless your meter is a Version “B”. To select a display color, follow these steps:
30
1.
Press the MENU button until the meter shows COLR.
2.
Press the ∂TARE button. The meter shows one of the following: •
GRN
•
RED
•
AMBR
3.
Press the ßMAX button to scroll between available choices.
4.
Press the MENU button to store your choice. The meter momentarily shows STRD , followed by the next menu S1.CF (Setpoint 1 Configuration). Or you can press the RESET button to abort and go back to the RD.CF menu.
Configuring The Meter
4
4.6 USING SETPOINT 1 CONFIGURATION S1.CF Note ☞
Refer to Table 6-1 for a summary list of menu configuration.
Setpoint 1 Configuration S1.CF is not active unless your meter has dual relay output capabilities. The LED's will display whether the S1.CF is active or not. You may use Setpoint 1 Configuration S1.CF for the following: •
To set the setpoint's active band above or below your chosen value
•
To select whether the setpoint operation is latched or unlatched
4.6.1 Setting Setpoint 1's Active Band 1.
Press the MENU button until the meter shows S1.CF.
2.
Press the ∂TARE button. The meter shows one of the following: • S.1=A (Active above the setpoint) (Default) • S.1=B (Active below the setpoint)
3.
Press the ßMAX button to toggle between available choices.
4.
Press the ∂TARE button to select if Setpoint 1 is latched or unlatched or press the MENU button to store your selection.
4.6.2 Selecting if Setpoint 1 is Latched or Unlatched 1.
Press the MENU button until S1.CF displays, then press the ∂TARE button twice. or Press the ∂TARE button from S.1 . The meter shows one of the following: •
S.2=U Setpoint 1 to be unlatched (Default)
•
S.2=L Setpoint 1 to be latched
2.
Press the ßMAX button to toggle between available choices.
3.
Press the MENU button to store your selection(s). The meter momentarily shows STRD , followed by S2.CF (Setpoint 2 Configuration). 31
4
Configuring The Meter 4.7 USING SETPOINT 2 CONFIGURATION S2.CF Note ☞
Refer to Table 6-1 for a summary list of menu configuration.
Setpoint 2 Configuration S2.CF is not active unless your meter has dual relay output capabilities. The LED's will display whether the S2.CF is active or not. You may use Setpoint 2 Configuration S2.CF for the following: • To set the setpoint's active band above or below your chosen value • To select whether the setpoint operation is latched or unlatched
4.7.1 Setting Setpoint 2's Active Band 1.
Press the MENU button until the meter shows S2.CF .
2.
Press the ∂TARE button. The meter shows one of the following: •
S.1=A (Active above the setpoint) (Default)
•
S.1=B (Active below the setpoint)
3.
Press the ßMAX button to toggle between available choices.
4.
Press the ∂TARE button to select if Setpoint 2 is latched or unlatched or press the MENU button to store your selection and enter S1.DB (Setpoint 1 Deadband)
4.7.2 Selecting if Setpoint 2 is Latched or Unlatched 1.
Press the MENU button until S2.CF displays, then press the ∂TARE button twice. or Press the ∂TARE button from S.1 . The meter shows one of the following:
32
•
S.2=U Setpoint 2 to be unlatched (Default)
•
S.2=L Setpoint 2 to be latched
2.
Press the ßMAX button to toggle between available choices.
3.
Press the MENU button to store your selection(s). The meter momentarily shows STRD , followed by S1.DB (Setpoint 1 Deadband).
Configuring The Meter
4
4.8 SETTING THE SETPOINT 1 DEADBAND S1.DB Note ☞
Refer to Table 6-1 for a summary list of menu configuration.
Setpoint 1 Deadband S1.DB is not active unless your meter has dual relay output capabilities. The LED's will display whether the S1.DB is active or not. The Setpoint 1 Default deadband is 0003. To change the deadband (hysteresis) of Setpoint 1, follow these steps: 1.
Press the MENU button until the meter shows S1.DB .
2.
Press the ∂TARE button. The meter shows the last previously stored 4-digit number (0000 through 9999) with flashing 4th digit.
3.
Press the ßMAX button to change the value of the flashing digit. If you continue to press the ßMAX button, the flashing digit's value continues to change.
4.
Press the ∂TARE button to scroll to the next digit.
5.
Press the MENU button to store your selection. The meter momentarily shows STRD , followed by S2.DB (Setpoint 2 Deadband).
4.9 SETTING THE SETPOINT 2 DEADBAND S2.DB Note ☞
Refer to Table 6-1 for a summary list of menu configuration.
Setpoint 2 Deadband S2.DB is not active unless your meter has dual relay output capabilities. The LED's will display whether the S2.DB is active or not. The Setpoint 2 default deadband is 0003. To change the deadband (hysteresis) of Setpoint 2, follow these steps: 1.
Press the MENU button until the meter shows S2.DB .
2.
Press the ∂TARE button. The meter shows the last previously stored 4-digit number (0000 through 9999) with flashing 4th digit.
3.
Press the ßMAX button to change the value of the flashing digit. If you continue to press the ßMAX button, the flashing digit's value continues to change.
4.
Press the ∂TARE button to scroll to the next digit.
5.
Press the MENU button to store your selection. The meter momentarily shows STRD , followed by OT.CF (Output Configuration) if you have analog output capabilities. 33
4
Configuring The Meter Figure 4-1. Alarm Example
SETPOINT SIGNAL LEVEL ON
OFF
ON
OFF
ACTIVE BELOW
ON
OFF
ACTIVE ABOVE
3 DEADBAND
SETPOINT 3
SIGNAL LEVEL OFF
ON
ON
OFF
ON
OFF
ON
ACTIVE BELOW WITH DEADBAND 3
ACTIVE ABOVE WITH DEADBAND 3
NOTE: DEADBAND WORKS AS HYSTERISIS
SETPOINT SIGNAL LEVEL
ON
ON
ON
ACTIVE BELOW LATCHED
Note ☞
OFF
ON
ACTIVE ABOVE LATCHED
To reset latched alarms you must: 1. Input a signal OUT of the alarm zone 2. Then press SETPTS and then, RESET button
34
ON
Configuring The Meter
4
4.10 USING OUTPUT CONFIGURATION OT.CF Note ☞
Refer to Table 6-1 for a summary list of menu configuration.
Output Configuration OT.CF is not active unless your meter has analog output capabilities. The menu will display whether analog output is present or not. Analog output must be ordered at the time of purchase. Use Output Configuration OT.CF to select the following: •
To enable or disable the analog output
•
To select if the analog output is current or voltage
•
To select if the analog output is regular or proportional
4.10.1 Enabling or Disabling the Analog Output To enable or disable the analog output, follow these steps: 1.
Press the MENU button until the meter shows OT.CF.
2.
Press the ∂TARE button. The meter shows one of the following: •
O.1=E (Analog output enabled) (Default)
•
O.1=D (Analog output disabled)
3.
Press the ßMAX button to toggle between available choices.
4.
Press the ∂TARE button to select analog output as current or voltage or press the MENU button to store your selection and enter OT.S.O (Output Scale and Offset).
35
4
Configuring The Meter 4.10.2 Selecting Analog Output as Current or Voltage 1.
Press the MENU button until it shows OT.CF, then press the ∂TARE button twice. or Press the ∂TARE button from 0.1 . The meter shows one of the following: •
O.2=C (Analog output = current) (Default)
•
O.2=V (Analog output = voltage)
2.
Press the ßMAX button to toggle between available choices.
3.
Press the ∂TARE button to select analog output or proportional control or press the MENU button to store your selection and enter OT.S.O (Output Scale and Offset).
4.10.3 Selecting Analog Output or Proportional Control Note ☞
Use this section to select if the meter will transmit an analog signal proportional to the display readings, or proportional to the error signal between the display reading and Setpoint 1.
Proportional Control Analog Option is not available for models without Relay Option. 1.
2.
36
Press the MENU button until it shows OT.CF , then press the ∂TARE button twice. or Press the ∂TARE button from 0.2 . The meter shows one of the following: •
O.3=A (Analog output is regular) (Default)
•
O.3=P (Analog output is proportional)
Press the ßMAX button to toggle between available choices.
Configuring The Meter
4
4.10.3 Selecting Analog Output or Proportional Control (Continued) 3a.
If you select O.3=A , press the MENU button to store your selection. The meter momentarily shows STRD , followed by OT.S.O (Output Scale and Offset).
3b.
If you select O.3=P , press the ∂TARE button. The meter shows one of the following: • •
O.4=D O.4=R
(Proportional analog output is DIRECT ACTING) (Proportional analog output is REVERSE ACTING).
4.
Press the ßMAX button to toggle between available choices.
5.
Press the MENU button to store your selections. The meter momentarily shows STRD , followed by P.BND (Proportional Band). Additionally, if you select O.2=V (Analog output to be voltage), press the ∂TARE button. One of the following displays: • •
O.5=F O.5=H
(Proportional 0-10 V analog output) (Proportional 0-5 V analog output).
6.
Press the ßMAX button to toggle between available choices.
7.
Press the MENU button to store your selections. The meter momentarily shows STRD , followed by P.BND (Proportional Band).
37
4
Configuring The Meter 4.11 SELECTING PROPORTIONAL BAND P.BND Proportional Band P.BND is not active unless your meter has analog output and relay capabilities. The menu will display whether analog output is present or not. •
A proportional controller's output is linearly proportional to the change of the error signal, whenever the signal is within 2 prescribed values (Proportional Band).
•
There are three (3) points of interest on the proportional controller transfer curve.
•
The first is the magnitude of the error signal that drives the controller to “full on” (e.g. 20 mA out for 4-20 mA). CONTROLLER OUTPUT
100%
OFFSET
Zero error
0%
ERROR=READING SETPOINT MIN
MAX
PROPORTIONAL BAND
4mA
20mA
Figure 4-2. Controller Output •
The second point of interest is the magnitude of the error signal that drives the controller output to “full off” (e.g. 4 mA out on 4-20 mA). These two (2) points need not be equally spaced on either side of the zero error point.
•
The third is the factor "Offset" and it is the output value of the controller which causes zero error.
Note ☞
38
The above example illustrates the parameters for the 4-20 mA analog out, likewise, analog voltage output will have these (3) points of interest.
Configuring The Meter
4
4.11 SELECTING PROPORTIONAL BAND P.BND (Continued) If A is the controller gain then, Proportional Band = CONTROLLER OUT =
Max. out - Min. out A A • ERROR + OFFSET
To select the proportional band for your proportional controller, follow these steps: 1.
Press the MENU button until the meter shows P.BND .
If P.BND menu doesn’t show, set 0.3=P on Menu OT.CF .
Note ☞
Remember to press ∂TARE when OT.CF is displayed until 0.3=A , then press ßMAX, unit will show 0.3=P . Pressing the MENU button will store the selection.
2.
Press the ∂TARE button. The meter shows last previously stored 4-digit number (0000 through 9999) with flashing 4th digit.
3.
Press the ßMAX button to change the value of the flashing digit. If you continue to press the ßMAX button, the flashing digit's value continues to change.
4.
Press the ∂TARE button to scroll to the next digit.
5.
Press the MENU button to store your selection. The meter shows STRD , followed by M.RST (Manual Reset).
39
4
Configuring The Meter 4.12 USING MANUAL RESET M.RST Note ☞
Refer to Table 6-1 for a summary list of menu configuration.
Manual Reset M.RST is not active unless your meter has analog output and relay capabilities. The menu will display whether analog output is present or not. This feature allows you to offset the error that may occur with your setpoint. In order to determine the amount of error, you must compare your display value to the Setpoint 1 value. The difference between these two values (display - Setpoint 1) is the amount of error that you may want to enter into Manual Reset M.RST . The value of M.RST must be less than P.BND /2. Larger values will not be accepted and the meter will display ER4 (flashing). 1.
Press the MENU button until M.RST displays. Note ☞
40
This menu M.RST and P.BND will show up if 0.3=P on OT.CF.
2.
Press the ∂TARE button. The meter shows the last previously stored 4-digit number (-1999 through 9999) with flashing 4th digit.
3.
Press the ßMAX button to change the value of the flashing digit. If you continue to press the ßMAX button, the flashing digit's value continues to change.
4.
Press the ∂TARE button to scroll to the next digit.
5.
Press the MENU button to store your selection. STRD momentarily displays, followed by RST (Reset).
Configuring The Meter
4
4.13 USING OUTPUT SCALE AND OFFSET OT.S.O Note ☞
Refer to Table 6-1 for a summary list of menu configuration.
Output Scale and Offset OT.S.O is not active unless your meter has analog output capabilities. The menu will display whether analog output is present or not. Output Scale and Offset OT.S.O scales your analog output to be equal to the meter's display and/or any engineering units you require. You may scale the output for direct (4-20 mA, 0-10 V, etc) or reverse acting (20-4 mA, 10-0 V, etc). 1.
Press the MENU button until OT.S.O displays.
2.
Press the ∂TARE button. RD1 (Read 1) displays. Note ☞
This is your first point of display reading.
3.
Press the ∂TARE button again. The meter shows the last previously stored 4-digit number (-1999 through 9999) with flashing 4th digit.
4.
Press the ßMAX button to change the digits.
5.
Press the ∂TARE button to scroll to the next digit.
6.
Press the MENU button to store your selection. OUT.1 (Output 1) displays. Note ☞
7.
This starting analog signal corresponds to your Read 1 display.
Press the ∂TARE button. Selected output displays. Note ☞
If you select O.2=V for voltage, the maximum signal you may select is 10.00 for an 0-10 Vdc signal output. If you select O.2=C for current, the maximum signal you may select is 20.00.
8.
Press the ßMAX button to enter the Output 1 signal selection. If you continue to press the ßMAX button, the flashing digit's value continues to change.
9.
Press the ∂TARE button to scroll to the next digit.
10.
Press the MENU button to store your selection. RD!2 (Read 2) displays. Note ☞
This is your second point of display reading. 41
4
Configuring The Meter 4.13 USING OUTPUT SCALE AND OFFSET OT.S.O (Continued) 11.
Press the ∂TARE button. The meter shows last previously stored 4-digit number (-1999 through 9999) displays with flashing 4th digit.
12.
Press the ßMAX button to change the value of the flashing digit. If you continue to press the ßMAX button, the flashing digit's value continues to change.
13.
Press the ∂TARE button to scroll to the next digit.
14.
Press the MENU button to store your selection. The meter shows OUT.2 (Output 2). Note ☞
This analog signal should correspond to your Read 2 display. 15.
Note ☞
Press the ∂TARE button. The meter shows selected output.
If you select O.2=V for voltage, the maximum signal you may select is 10.00 for an 0-10 Vdc signal output. If you select O.2=C for current, the maximum signal you may select is 20.00 for a 0-20 or 4-20 mA DC signal output.
16.
Press the ßMAX button to change the value of the flashing digit. If you continue to press the ßMAX button, the flashing digit's value continues to change.
17.
Press the ∂TARE button to scroll to the next digit.
18.
Press the MENU button to store your selection. The meter momentarily shows STRD , followed by LK.CF (Lockout Configuration). WARNING: If the meter displays all flashing values on any item, the value has overflowed. Press the ßMAX button to start new values.
42
Configuring The Meter
4
4.14 USING LOCK OUT CONFIGURATION LK.CF Note ☞
Refer to Table 6-1 for a summary list of menu configuration.
Use Lock Out Configuration LK.CF for the following: •
To enable or disable setpoint changes
•
To enable or disable the RESET button in the Run Mode
•
To enable or disable displaying the meter’s firmware version.
4.14.1 Enabling or Disabling the RESET button in the Run Mode 1.
Press the MENU button until the meter shows LK.CF (after OT.S.O ).
2.
Press the ∂TARE button. The meter shows one of the following: • RS.=E To enable the RESET button in the Run Mode (Default) •
RS.=D To disable the RESET button in the Run Mode
3.
Press the ßMAX button to toggle between available choices.
4.
Press the MENU button to store the changes. The meter shows STRD if the new value is different otherwise the meter shows BRIT and returns to the Run Mode.
4.14.2 Enabling or Disabling SETPOINT Changes 1.
Press the MENU button until the meter shows LK.CF (after OT.S.O ).
2.
Press the ∂TARE button twice. The meter shows one of the following: •
SP.=E To enable setpoint changes (Default)
•
SP.=D To disable setpoint changes
3.
Press the ßMAX button to toggle between available choices.
4.
Press the MENU button to store the changes. The meter shows STRD if the new value is different otherwise the meter shows BRIT and returns to the Run Mode.
43
4
Configuring The Meter 4.14.3 SETPOINT Display Function: Firmware version or Setpoint value 1.
Press the MENU button until the meter shows LK.CF (after OT.S.O ).
2.
Press the ∂TARE button three times. The meter shows one of the following: • L.3=0 SETPTS button will display setpoint values. • L.3=1 SETPTS button will display the meter's firmware version.
3.
Press the ßMAX button to toggle between the choices above.
4.
Press the MENU button to store the changes. The meter shows STRD if the new value is different otherwise the meter shows BRIT and returns to the Run Mode.
Note ☞
If your meter does not have the relay option, setpoint menu items above will not be available and SETPTS button will always display the meter's firmware version. These units will have +OL (overload) or +OPN memory indicated by Alarm 1 & 2 LED displays. LEDs can be reset by pressing MENU then RESET button or by Power OFF then ON.
4.15 USING DISPLAY BRIGHTNESS CONFIGURATION Changing “Display Brightness” is not active unless your meter is a Version “B”.
44
1.
Press the MENU button until the meter shows BRIT (after LKCF).
2.
Press the ∂TARE button from BRIT. The meter shows one of the following: •
M.BrT Medium Brightness
•
L.BrT Low Brightness
•
H.BrT High Brightness (Default)
3.
Press the ßMAX button to toggle between available choices.
4.
Press the MENU button to store your selection. The meter momentarily shows STRD followed by STRD, RST, PROC, then measured value.
Display Messages
5
SECTION 5. DISPLAY MESSAGES Table 5-1. Display Messages MESSAGE PROC RST INPT DEC.P RD.S.O RD.CF COLR S1.CF S2.CF S1.DB S2.DB P.BND M.RST ER4 OT.CF OT.S.O LK.CF BRIT +OL -OL RS.OF +999 -1999 ER1 PEAK PK.RS T.RS SP.RS SP1 SP2 R.OV.S V.-8.8 RUN
DESCRIPTION Process Meter Hard (Power On) Reset Input Type Decimal Point Reading Scale and Offset Reading Configuration Display Color Setpoint 1 Configuration Setpoint 2 Configuration Setpoint 1 Deadband Setpoint 2 Deadband Proportional Band Manual Reset Manual Reset Error Output Configuration Output Scale and Offset Lock Out Configuration Display Brightness + Overload Signal - Overload Signal Resolution Overflow Value Overflow in Setpoint/Menu & Peak Routines Value Overflow in Setpoint/Menu & Peak Routines 2 Coordinate Format Programming Error Peak Value Peak Reset Tare Reset Reset Latched Alarms Setpoint 1 Value Setpoint 2 Value Resolution Over Scale Firmware Version (where 8 is 0 ~ 9) Operating Mode
45
6
Menu Configuration Displays SECTION 6. MENU CONFIGURATION DISPLAYS Not all menu items display on standard meters.
Table 6-1. Menu Configuration Displays (Defaults in Bold and Italics) ∂TARE
MENU
100M ±50M 10V ±5V 0-20 (Default)
Show current decimal point position
FFFF (Default) F.FFF FF.FF FFF.F
INPT DEC.P Reading Scale & Offset
RD.S.O
1
ßMAX
Show input choices:
Shows IN1 Shows prior value entered and flashing digit. Scrolls to the next digit. • If ∂TARE is pressed, actual input
Changes the value of the flashing digit
is shown and can not be changed with ßMAX. • If ßMAX is pressed, unit can scroll through digits with ∂TARE.
2
Enter new value and show RD1
prior value entered and flashing digit. 3 Shows Scrolls to the next digit.
Changes the value of the flashing digit
4
Enter new value and show IN2
prior value entered and flashing digit. 5 Shows Scrolls to the next digit.
Changes the value of the flashing digit
• If ∂TARE is pressed, actual input
is shown and can not be changed with ßMAX. • If ßMAX is pressed, unit can scroll through digits with ∂TARE.
6
46
Enter new value and show RD2
prior value entered and flashing digit. 7 Shows Scrolls to the next digit.
Changes the value of the flashing digit
Menu Configuration Displays
6
SECTION 6. MENU CONFIGURATION DISPLAYS (Continued) Table 6-1. Menu Configuration Displays (Continued) (Defaults in Bold and Italics) MENU
∂TARE
ßMAX
Reading Configuration
R.1=
R.1=T (Tare enabled) R.1=N (Tare disabled)
RD.CF
R.2=
R.2=0 (1 µV resolution for unipolar & 5 µV resolution for bipolar) R.2=1 (2 µV resolution for unipolar & 10 µV resolution for bipolar) R.2=2 (3 µV resolution for unipolar & 15 µV resolution for bipolar) R.2=3 (5 µV resolution for unipolar & 25 µV resolution for bipolar) R.2=4 (10 µV resolution for unipolar & 25 µV resolution for bipolar) Note: 3 µV resolution means if your input is 0-30 mV, at 30 mV the display shows 9999.
R.3=
Display Color Selection
Show input choices:
COLR
(Green) (Red)
S.1=
S.1=A (Active above) S.1=B (Active below)
S.2=
S.2=U (Unlatched) S.2=L (Latched)
S.1=
S.1=A (Active above) S.1=B (Active below)
S.2=
S.2=U (Unlatched) S.2=L (Latched)
Setpoint 2 Configuration
S2.CF
GRN RED
AMBR (Amber)
Setpoint 1 Configuration
S1.CF
R.3=F (Filtered value) R.3=U (Unfiltered value)
47
6
Menu Configuration Displays SECTION 6. MENU CONFIGURATION DISPLAYS (Continued) Table 6-1. Menu Configuration Displays (Continued) (Defaults in Bold and Italics) ∂TARE
MENU Setpoint 1 Deadband
ßMAX
Press to scroll to the next digit to the right
Press to change the value of the flashing digit
Press to scroll to the next digit to the right
Press to change the value of the flashing digit
S1.DB Setpoint 2 Deadband
S2.DB Output Configuration
OT.CF
0.1 =
0.1=E (Analog output is enabled) 0.1=D (Analog output is disabled)
0.2 =
0.2=C (Analog output is current) 0.2=V (Analog output is voltage)
0.3 =
0.3=A (Regular analog output) 0.3=P (Proportional analog output)*
0.4 = shown if menu 0.3=P
0.4=D (Proportional analog is direct acting) 0.4=R (Proportional analog is reverse acting)
0.5 = shown if menu 0.2=V
0.5=F (Analog output is 0-10 Vdc) 0.5=H (Analog output is 0-5 Vdc)
Note ☞
* If you select 0.2V and 0.3=P , you may select your analog output to be 0-10 V or 0-5 V by accessing submenu 0.5=F or 0.5=H. * If 0.3=P, you may select your proportional output analog to be: • Direct Acting 0.4=D: 4-20 mA, 0-5 V, 0-10 V • Reverse Acting 0.4=R: 20-4 mA, 5 V-0 V, 10 V-0 V
48
Menu Configuration Displays
6
SECTION 6. MENU CONFIGURATION DISPLAYS (Continued) Table 6-1. Menu Configuration Displays (Continued) (Defaults in Bold and Italics) ∂TARE
MENU
P.BND
Proportional Band shown menu if 0.3=P
M.RST
Manual Reset shown menu if 0.3=P
Shows prior value entered. Scrolls to the next digit to the right.
Changes the value of the flashing digit
Shows prior value entered. Scrolls to the next digit to the right.
Changes the value of the flashing digit
Output Scale & Offset
OT.S.O
ßMAX
1
Shows RD1 Shows prior value entered and flashing digit. Scrolls to the next digit.
Changes the value of the flashing digit
(Shown if O.3=A in Output Configuration Menu OT.CF )
2
Enter new value and show OUT1
3
Shows prior value entered and flashing digit. Scrolls to the next digit.
Changes the value of the flashing digit
4
Enter new value and show RD2
5
Shows prior value entered and flashing digit. Scrolls to the next digit.
Changes the value of the flashing digit
6
Enter new value and show OUT2
7
Shows prior value entered and flashing digit. Scrolls to the next digit.
Changes the value of the flashing digit
49
6
Menu Configuration Displays SECTION 6. MENU CONFIGURATION DISPLAYS (Continued) Table 6-1. Menu Configuration Displays (Continued) (Defaults in Bold and Italics) MENU
∂TARE
Lock Out Configuration
RS=
RS=E (Enable RESET button in the Run Mode) RS=D (Disable RESET button in the Run Mode)
ßMAX
LK.CF
SP=
SP=E (Enable setpoint changes) SP=D (Disable setpoint changes)
L3=
L3=0 (SETPTS button display setpoint values) L3=1 (SETPTS button display firmware version V.-8.8 where 8 is 0~9) M.BrT (Medium Brightness)
Brightness Configuration
BRIT
L.BrT (Low Brightness) H.BrT (High Brightness)
Displays Display
Table 6-2. Run Mode Table 6-2. Run Mode Displays ∂TARE
PEAK
T.RST SP.RS 50
Press to activate.
ßMAX Displays the peak reading and must be pressed again to return to the normal operating mode without resetting.
RESET Reset the peak reading when in this mode.
Description Peak Reading Displays the highest reading since last reset.
Will reset your Tare Reset tare when viewing this function. Reset Latched Alarms Pressing the RESET button resets your latched alarms.
Setpoint Configuration Displays
7
SECTION 7. SETPOINT CONFIGURATION DISPLAYS Table 7-1. Setpoint Configuration Displays
MENU
SP1 SP2
∂TARE
ßMAX
Description SETPOINT 1
Press to scroll to the next digit to the right
Press to change the value of the flashing digit
Select from -1999 through 9999 SETPOINT 2
Press to scroll to the next digit to the right
Press to change the value of the flashing digit
Select from -1999 through 9999
51
8
Specifications SECTION 8. SPECIFICATIONS SIGNAL INPUT Input Ranges:
0-100 mV, ± 50 mV, 0-10 V, ± 5 V, 0-20 mA, 4-20 mA
Isolation:
Dielectric strength to 2500V transient per 3mm spacing based on EN 61010 for 260Vrms or DC working voltage
Noise Rejection:
Normal Mode Rejection (NMR) = 60 dB Common Mode Rejection (CMR) = 120 dB
Resistance:
100 Meg ohms for 100 mV or ±50 mV input range 1 Meg ohm for 10 V or +5 V input range 5 ohms for 20 mA current input range
“Big” Display:
4-digit, three color programmable 9-segment, LED 21 mm (0.83")
( -1.9.9.9 ~ 9.9.9.9)
Symbol: Standard Display:
4-digit, 14-segment LED, 13.8 mm (0.54”)
( -1.9.9.9. ~ 9.9.9.9.)
Symbol: ANALOG TO DIGITAL Technique:
Dual slope
Internal resolution:
15 bits
Read Rate:
3/sec Polarity Automatic
ACCURACY AT 25°C Max Error Strain/Process:
±0.03% of reading, ±1 count
Span Tempco:
50 ppm/°C
Step Response:
1 sec
Warm Up to Rated Accuracy: 30 min
52
Excitation Voltage:
AC power units 24 V @ 25 mA, 12 V @ 50 mA, 10 V @ 120 mA, 5 V @ 60 mA
Load Regulation:
1.1%
Line Regulation:
0.02% per Vac
DC power units Refer to Table 8-1 Color chart for DC Output Excitation
Specifications
8
SECTION 8. SPECIFICATIONS (Continued) ALARM OUTPUTS (if applicable) 2 Form "C" on/off relays. Configurable for latched and unlatched by software. Max current: 5 AMPS, resistive load Max voltage: 250 Vac or 30 Vdc
ANALOG OUTPUT (if applicable) Signal Type:
Current or voltage
Signal Level:
Current: 10 V max compliance at 20 mA output Voltage: 20 mA max for 0-10 V output
Function:
May be assigned to a display range or proportional control output with Setpoint #1 when used as a control output.
Linearity:
0.2%
Step Response Time:
2 - 3 seconds to 99% of the final value
ISOLATED ANALOG OUTPUT (TB5, if applicable) Same as non-isolated analog output except isolated. Signal Type:
Current or voltage
Signal Level:
Current: 10 V max compliance at 20 mA output Voltage: 20 mA max for 0-10 V output
Function:
May be assigned to a display range or proportional control output with Setpoint #1 when used as a control output.
Linearity:
0.2%
Step Response Time:
2 - 3 seconds to 99% of the final value
Isolation
130 Vrms working voltage, 1000 V/60sec Dielectric test
Note ☞
• Only one analog output is available on each unit and it must be factory installed.
53
8
Specifications SECTION 8. SPECIFICATIONS (Continued) INPUT POWER INFORMATION ~ AC units
DC units
115/230 V~(AC) + 10%, 50/60 Hz 9.5 W max, power consumption (Non-Isolated Analog Out) 11.0 W max, power consumption (Isolated Analog Out) 10-32 Vdc or 26-56 Vdc, 8W (Refer to Table 8-1 below)
External Fuse Required:
IEC 127-2/III Power
Fuse
115 V
125 mA @ 250 (T)
230 V
63 mA @ 250 (T)
UL 248-14 (Listed Fuse) Power
Fuse
115 V
175 mA @ 250 V Slow-Blow
230 V
80 mA @ 250 V Slow-Blow
ENVIRONMENT Operating temperature:
0° to 50°C (32° to 122°F)
Storage temperature:
-40° to 85°C (-40° to 185° F)
Relative humidity:
90% at 40°C (non-condensing)
MECHANICAL
54
Panel cutout:
1/8 DIN 3.62 x 1.78" (45 x 92mm)
Weight:
1.27 lb (575 g)
Case material:
Polycarbonate, 94 V-O UL rated
Protection:
NEMA-4/Type 4 Front Bezel
Specifications
8
SECTION 8. SPECIFICATIONS (Continued) Table 8-1. COLOR CHART FOR DC POWER COLOR
HIGH BRIGHTNESS
MEDIUM & LOW BRIGHTNESS
RED
Sensor Excitation: 24 V @ 25 mA, 12 V, 10 V, 5 V @ 35 mA Max Analog Output: Non-Isolated option only
GREEN
Warning:
Any combination of Sensor Excitation and Analog Output 24 V @ 25 mA, 12 V @ 35 mA Max 10 V @ 35 mA Max 5 V @ 35 mA Max
AMBER
Note ☞
Analog Output: • Do not use Internal Excitation. Non-Isolated options or Use External Excitation. Isolated Analog option • Do not use Isolated Analog Output. Use Non-Isolated Analog Ouput.
HIGH/LOW Brightness and AMBER are only available on Version “B” meters. Standard display meters are MEDIUM Brightness.
55
8
Specifications SECTION 8. SPECIFICATIONS (Continued)
48,0 (1.89)
96,0 (3.78) FRONT BEZEL
20,3 (.80)
151,4 (5.96)
RETAINER
CASE
REAR COVER
SIDE VIEW
TOP VIEW
PANEL THICKNESS 6,4 (.25) MAX 0,8 (.03) MIN
1,5 R(.06) 4 PLCS
45,00 + 0,61/-0,00 (1.772 + .024/–.000)
92,00 + 0,81/–0,00 (3.622 + .032/–.000) NOTE: Dimensions in Millimeters (Inches)
Figure 8-1 Meter Dimensions/ Panel Cutout
56
Factory Preset Values
9
SECTION 9. FACTORY PRESET VALUES Table 9-1. Factory Preset Values MENU ITEM INPT DEC.P RD.S.O
FACTORY PRESET VALUES Input Type: 0-20 (0-20 mA dc input) Decimal Point Position: FFFF Reading Scale and Offset: 4-20 mA dc = 0-1000 RD.CF Reading Configuration: R.1=T (Tare enabled) R.2=4 (10 µV resolution for unipolar & 25 µV resolution for bipolar) R.3=F (Filtered value) COLR Normal Color Display: RED or GRN (Note: depending how unit was ordered) S1.CF Setpoint 1 Configuration: S.1=A (Setpoint is active above) S.2=U (Setpoint is unlatched) S2.CF Setpoint 2 Configuration: S.1=A (Setpoint is active above) S.2=U (Setpoint is unlatched) S1.DB Setpoint 1 Deadband: 0003 S2.DB Setpoint 2 Deadband: 0003 OT.CF Output Configuration: O.1=E (Analog output is enabled) O.2=C (Analog output is current) O.3=A (Analog output follows the display value) OT.S.O Output Scale and Offset: 0-1000 = 4-20 mA dc LK.CF Lock Out Configuration RS=E (Enable the RESET button in the Run Mode) SP=E (Enable setpoint changes) L3=0 (Display setpoint values) BRIT H.BrT (Brightness Level) SP1 Setpoint 1 Value: 0000 SP2 Setpoint 2 Value: 0000 Sensor Excitation Output: 12 Vdc
57
CE APPROVALS INFORMATION This product conforms to the EMC directive 89/336/EEC amended by 93/68/EEC, and with the European Low Voltage Directive 72/23/EEC. Electrical Safety EN61010-1:2001 Safety requirements for electrical equipment for measurement, control and laboratory. Double Insulation Pollution Degree 2 Dielectric withstand Test per 1 min • Power to Input/Output: 2300 Vac (3250 Vdc) • Power to Input/Output: 500 Vac (720 Vdc) (Low Voltage dc Power Option*)
• Power to Relays Output: 2300 Vac (3250 Vdc) • Relay 1 to Relay 2: 2300 Vac (3250 Vdc) • Isolated Analog to Inputs: 1000 Vac (1420 Vdc) • Analog to Inputs: No Isolation Measurement Category I Category I are measurements performed on circuits not directly connected to the Mains Supply (power). Maximum Line-to-Neutral working voltage is 50 Vac/dc. This unit should not be used in Measurement Categories II, III, IV. Transients Overvoltage Surge (1.2 / 50uS pulse) • Input Power: 2500 V • Input Power: 500 V (Low Voltage dc Power Option*)
• Isolated Analog: 500 V • Input/Output Signals: 500 V Note: *Units configured for external low power dc voltage, 10-32 Vdc (Basic Insulation) EMC EN61326:1997 + and A1:1998 + A2:2001 Immunity and Emissions requirements for electrical equipment for measurement, control and laboratory. • EMC Emissions Table 4, Class B of EN61326 • EMC Immunity** Table 1 of EN61326 Note: **I/O signal and control lines require shielded cables and these cables must be located on conductive cable trays or in conduits. Furthermore, the length of these cables should not exceed 30 meters Refer to the EMC and Safety installation considerations (Guidelines) of this manual for additional information.
58
NOTES
59
NOTES
60
Warranty/Disclaimer NEWPORT Electronics, Inc. warrants this unit to be free of defects in materials and workmanship for a period of one (1) year from the date of purchase. In addition to NEWPORT’s standard warranty period, NEWPORT Electronics will extend the warranty period for four (4) additional years if the warranty card enclosed with each instrument is returned to NEWPORT. If the unit should malfunction, it must be returned to the factory for evaluation. NEWPORT’s Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by NEWPORT, if the unit is found to be defective it will be repaired or replaced at no charge. NEWPORT’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of being damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of NEWPORT’s control. Components which wear are not warranted, including but not limited to contact points, fuses, and triacs. NEWPORT is pleased to offer suggestions on the use of its various products. However, NEWPORT neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by NEWPORT, either verbal or written. NEWPORT warrants only that the parts manufactured by it will be as specified and free of defects. NEWPORT MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive and the total liability of NEWPORT with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall NEWPORT be liable for consequential, incidental or special damages. CONDITIONS: Equipment sold by NEWPORT is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, or used on humans, or misused in any way, NEWPORT assumes no responsibility as set forth in our basic WARRANTY / DISCLAIMER language, and additionally purchaser will indemnify NEWPORT and hold NEWPORT harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.
Return Requests/Inquiries Direct all warranty and repair requests/inquiries to the NEWPORT Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO NEWPORT, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM NEWPORT’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence. The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit. FOR WARRANTY RETURNS, please have the following information available BEFORE contacting NEWPORT:
FOR NON-WARRANTY REPAIRS, consult NEWPORT for current repair charges. Have the following information available BEFORE contacting NEWPORT:
1.
1.
2. 3.
P.O. number under which the product was PURCHASED, Model and serial number of the product under warranty, and Repair instructions and/or specific problems relative to the product.
2. 3.
P.O. number to cover the COST of the repair, Model and serial number of product, and Repair instructions and/or specific problems relative to the product.
NEWPORT’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering. NEWPORT is a registered trademark of NEWPORT Electronics, Inc. © Copyright 2006 NEWPORT Electronics, Inc. All rights reserved. This document may not be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without prior written consent of NEWPORT Electronics, Inc.
For immediate technical or application assistance please call:
Newport Electronics, Inc. 2229 South Yale Street • Santa Ana, CA • 92704 • U.S.A. TEL: (714) 540-4914 • FAX: (203) 968-7311 Toll Free: 1-800-639-7678 • www.newportUS.com • e-mail:
[email protected] ISO 9001 Certified
Newport Technologies, Inc. 976 Bergar • Laval (Quebec) • H7L 5A1 • Canada TEL: (514) 335-3183 • FAX: (514) 856-6886 Toll Free: 1-800-639-7678 • www.newport.ca • e-mail:
[email protected]
Newport Electronics, Ltd. One Omega Drive • River Bend Technology Centre Northbank, Irlam • Manchester M44 5BD • United Kingdom Tel: +44 161 777 6611 • FAX: +44 161 777 6622 Toll Free: 0800 488 488 • www.newportuk.co.uk • e-mail:
[email protected]
Newport Electronics spol s.r.o. Frystatska 184, 733 01 Karviná • Czech Republic TEL: +420 59 6311899 • FAX: +420 59 6311114 Toll Free: 0800-1-66342 • www.newport.cz • e-mail:
[email protected]
Newport Electronics GmbH Daimlerstrasse 26 • D-75392 Deckenpfronn • Germany TEL: 49 7056 9398-0 • FAX: 49 7056 9398-29 Toll Free: 0800 / 6397678 • www.newport.de • e-mail:
[email protected]
Mexico and Latin America FAX: 001 (203) 359-7807 En Español: 001 (203) 359-7803
M3597/N/1006