LAPORAN PROGRAM UJI BANDING (PROFICIENCY TEST) ANTAR LABORATORIUM SKEMA KHUSUS SEMESTER I-2015
Bahan Uji: AIR MINUM DALAM KEMASAN I (AMDK I) Ref. No. 005/UR/BMD/2015
PARAMETER UJI: Nitrat, Sulfat, dan Klorida
BMD Laboratory–Provider of Proficiency Testing Program Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email:
[email protected]
FR.06 Laporan Program Uji Banding (Final Report) Ed./Rev./Date: 01/02/16-04-2015
This is an official electronic document and secured by BMD Laboratory ©BMD Laboratory 2015
BMD Laboratory–Provider of Proficiency Testing Program Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email:
[email protected]
KATA PENGANTAR Puji syukur kiranya kita panjatkan kepada Alloh SWT yang berkenan melimpahkan karunia-Nya sehingga ‘Program Uji Banding Antar Laboratorium (Proficiency Test) Skema Khusus Semester I-2015’ untuk bahan uji Air Minum dalam Kemasan I (AMDK I) dengan parameter uji nitrat (NO3-), sulfat (SO42-), dan klorida (Cl-) telah berhasil dilaksanakan dan diselesaikan sampai dengan diterbitkannya laporan ini. Program uji banding antar laboratorium merupakan salah satu instrumen yang sangat penting sebagai sarana jaminan mutu eksternal (external quality control) bagi laboratorium pengujian dan kalibrasi sesuai dengan persyaratan standard SNI ISO/IEC 17025:2008. Dengan mengikuti program uji banding, laboratorium dapat melakukan evaluasi kinerja pengujian yang dilakukan dalam rangka melakukan peningkatan secara berkesinambungan (continuous improvement) untuk memberikan layanan terbaik kepada pelanggan. Program skema khusus kali ini diikuti oleh 11 (sebelas) laboratorium dimana 10 (sepuluh) laboratorium diantaranya terakreditasi SNI ISO/IEC 17025:2008 oleh Komite Akreditas Nasional (KAN). Seluruh data dapat diolah dengan baik menggunakan Robust Statistic Algoritma A Mean sesuai ISO 13528:2005(E) dengan terlebih dahulu dilakukan evaluasi data menggunakan perangkat lunak Distribution AnalyzerTM untuk memastikan data telah memenuhi distribusi normal. Dalam program uji banding ini masih digunakan nilai konsensus (consensus value) dari laboratorium peserta dalam melakukan evaluasi, sehingga tidak memuaskan dalam ketertelusuran (traceability) secara metrologi. Akan tetapi BMD Laboratory meyakini bahwa hal tersebut tidak mengurangi manfaat dari pelaksanaan program uji banding ini. Untuk itu BMD Laboratory akan terus mengupayakan perbaikan demi perbaikan guna menghadirkan program uji banding yang semakin bermanfaat bagi laboratorium-laboratorium di Indonesia. BMD Laboratory sebagai provider uji banding antar laboratorium menyampaikan apresiasi setinggi-tingginya kepada seluruh pihak yang telah membantu dan menyukseskan kegiatan ini. Ref. No. 005/UR/BMD/2015 FR.06 Laporan Program Uji Banding (Final Report) Ed./Rev./Date: 01/02/16-04-2015
www.ujiprofisiensi.com
Page i of iii
BMD Laboratory–Provider of Proficiency Testing Program Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email:
[email protected]
Khususnya kepada seluruh laboratorium peserta atas partisipasi aktifnya mengikuti program uji banding AMDK I. Ucapan terimakasih juga disampaikan kepada tim ahli yang telah bekerja keras mengolah data dan menyusun laporan ini. Akhirnya, BMD Laboratory menyampaikan permohonan maaf sekiranya terdapat kekurangan dalam penyelenggaraan program uji banding antar laboratorium. Kritik, saran, dan masukan yang membangun sangat kami harapkan untuk menghadirkan program uji banding yang semakin baik ke depannya. Semoga laporan ini dapat dimanfaatkan oleh laboratorium peserta dan pihak-pihak terkait.
Tangerang Selatan, 05 Juni 2015
Ir. Renaldo Moontri CEO BMD Group (Selaku Manajer Puncak BMD Laboratory)
Ref. No. 005/UR/BMD/2015 FR.06 Laporan Program Uji Banding (Final Report) Ed./Rev./Date: 01/02/16-04-2015
www.ujiprofisiensi.com
Page ii of iii
BMD Laboratory–Provider of Proficiency Testing Program Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email:
[email protected]
DAFTAR ISI KATA PENGANTAR .........................................................................................................................i DAFTAR ISI ................................................................................................................................... iii I.
PENDAHULUAN .................................................................................................................... 1
II.
PELAKSANAAN ..................................................................................................................... 1
III. STATISTIKA PENGOLAHAN DATA ......................................................................................... 2 III.1 Penilaian Laboratorium ............................................................................................... 2 III.2 Penetapan Nilai Evaluasi Uji Banding .......................................................................... 2 III.3 Penetapan SDPA .......................................................................................................... 2 IV. HASIL UJI BANDING ANTAR LABORATORIUM ...................................................................... 3 IV.1 Parameter Nitrat (NO3-) ............................................................................................. 4 IV.2 Parameter Sulfat (SO42-) ............................................................................................. 6 IV.3 Parameter Klorida (Cl-) ............................................................................................... 8 V.
PENUTUP ........................................................................................................................... 10
VI. REFERENSI .......................................................................................................................... 11 Annex A Rekapitulasi Hasil z-score Program Uji Banding ......................................................... 12 Annex B Rekapitulasi Bias Nilai Peserta Terhadap Nilai Konsensus ......................................... 13 Annex C Evaluasi Distribusi Normal Data Program Uji Banding ............................................... 15
Ref. No. 005/UR/BMD/2015 FR.06 Laporan Program Uji Banding (Final Report) Ed./Rev./Date: 01/02/16-04-2015
www.ujiprofisiensi.com
Page iii of iii
BMD Laboratory–Provider of Proficiency Testing Program Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email:
[email protected]
I. PENDAHULUAN Salah satu kebutuhan dasar manusia adalah air, terutama yang digunakan untuk minum. Kualitas kesehatan manusia salah satunya sangat ditentukan oleh kualitas air yang diminumnya. Oleh karena itu, air minum dalam kemasan (AMDK) merupakan salah satu komoditi yang diatur persyaratan mutunya sesuai SNI 01-3553-2006. Persyaratan mutu untuk parameter nitrat (NO3-), sulfat (SO42-), dan klorida (Cl-) seperti pada Tabel 1. Tabel 1. Persyaratan mutu air minum dalam kemasan (air mineral)[1] No.
Parameter
Satuan
Persyaratan
1.
Nitrat (NO3-)
mg/L
Maks. 45 mg/L
2.
Sulfat (SO42-)
mg/L
Maks. 200 mg/L
mg/L
Maks. 250 mg/L
3.
-
Klorida (Cl )
Salah satu jaminan mutu bagi laboratorium pengujian dan kalibrasi sesuai persyaratan dalam SNI ISO/IEC 17025:2008 adalah partisipasi dalam uji banding antar laboratorium atau uji profisiensi[2] sebagai jaminan mutu eksternal (external quality control). Untuk itu BMD Laboratory menyelenggarakan program uji banding antar laboratorium (proficiency test) untuk bahan uji air minum dalam kemasan I (AMDK I) dengan tujuan sebagai berikut: 1. Sebagai sarana jaminan mutu eksternal (external quality control) bagi laboratorium peserta khususnya laboratorium yang menerapkan standard SNI ISO/IEC 17025:2008. 2. Membantu laboratorium dalam mengevaluasi unjuk kerja pengujian kadar nitrat (NO3-), sulfat (SO42-), dan klorida (Cl-) untuk bahan uji air minum dalam kemasan. II. PELAKSANAAN Berikut adalah jadual pelaksanaan program uji banding antar laboratorium (proficiency test) yang diselenggarakan oleh BMD Laboratory untuk bahan uji air minum dalam kemasan I (AMDK I):
Ref. No. 005/UR/BMD/2015 FR.06 Laporan Program Uji Banding (Final Report) Ed./Rev./Date: 01/02/16-04-2015
www.ujiprofisiensi.com
Page 1 of 17
BMD Laboratory–Provider of Proficiency Testing Program Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email:
[email protected]
a. Penutupan pendaftaran: 30 Maret 2015 b. Pendistribusian bahan uji: 13 s.d. 30 April 2015 c. Pelaporan hasil analisis oleh laboratorium peserta: Maks. 21 Mei 2015 d. Pendistribusian laporan program uji banding oleh BMD Laboratory: Maks. 08 Juni 2015
III. STATISTIKA PENGOLAHAN DATA III.1 Penilaian Laboratorium Penilaian terhadap unjuk kerja laboratorium menggunakan z-score dengan tiga kriteria penilaian sebagai berikut: Untuk |z-score|≤2,0 dikategorikan memuaskan diberi lambang ‘OK’ Untuk 2,0<|z-score|<3,0 dikategorikan meragukan diberi lambang ‘$’ Untuk |z-score|≥3,0 dikategorikan kurang memuaskan diberi lambang ‘$$’ III.2 Penetapan Nilai Evaluasi Uji Banding Penetapan nilai evaluasi uji banding menggunakan dua pendekatan statistika yaitu: Robust Statistic Algoritma A Mean sesuai ISO 13528:2005(E)[3]. Nilai tersebut merupakan nilai konsensus dari peserta (consensus value) Seleksi data menggunakan perangkat lunak Distribution AnalyzerTM, Taylor Enterprises, Inc.[4] III.3 Penetapan SDPA SDPA (simpangan baku yang digunakan dalam evaluasi) ditetapkan menggunakan tiga pendekatan statistika yaitu: Robust Statistic Algoritma A Mean sesuai ISO 13528:2005(E) Simpangan Baku Horwitz[5] SDPA yang ditetapkan (Assigned SDPA)
Ref. No. 005/UR/BMD/2015 FR.06 Laporan Program Uji Banding (Final Report) Ed./Rev./Date: 01/02/16-04-2015
www.ujiprofisiensi.com
Page 2 of 17
BMD Laboratory–Provider of Proficiency Testing Program Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email:
[email protected]
IV. HASIL UJI BANDING ANTAR LABORATORIUM Sebanyak 11 (sebelas) laboratorium telah berpartisipasi dalam program uji banding antar laboratorium (proficiency test) yang diselenggarakan oleh BMD Laboratory untuk bahan uji air minum dalam kemasan I (AMDK I) dengan parameter uji nitrat (NO3-), sulfat (SO42-), dan klorida (Cl-). Adapun 10 (sepuluh) laboratorium diantaranya merupakan laboratorium terakreditasi SNI ISO/IEC 17025:2008 oleh Komite Akreditasi Nasional (KAN). Seluruh data hasil uji banding dapat diolah dengan baik menggunakan Robust Statistic Algoritma A Mean sesuai ISO 13528: 2005(E). Selain itu dilakukan juga evaluasi data menggunakan perangkat lunak Distribution AnalyzerTM guna memastikan bahwa data yang diolah mengikuti pola distribusi normal pada tingkat kepercayaan 95%. Nilai evaluasi uji banding (assigned value) merupakan nilai konsensus (consensus value) dari laboratorium peserta sehingga nilai tersebut bukan merupakan nilai acuan yang tertelusur secara metrologi ke sistem satuan internasional (SI). Tabel 2. Rekapitulasi Hasil Pengujian Laboratorium Peserta No.
Kode Lab.
Nitrat (NO3-) mg.L-1
Sulfat (SO42-) mg.L-1
Klorida (Cl-) mg.L-1
BMD-201-15
9.84 ± 0.06
4.68 ± 0.43
7.12 ± 0.81
Metode
SNI 01-3554-2006
SNI 01-3554-2006
SNI 01-3554-2006
BMD-202-15
-
5.64 ± 0.44
16.19 ± 0.24
Metode
-
SNI 6989.20:2009
SNI 6989.19:2009
BMD-204-15
3.46 ± 0.20
3.89 ± 2.43
6.92 ± 0.22
Metode
SNI 06-6989.79-2011
SNI 6989.20:2009
SNI 6989.19:2009
BMD-205-15
2.44
4.50
11.41
Metode
SNI 6989.79:2011
SNI 06-6989.20:2009
SNI 06-6989.19:2009
1.
2.
3.
4.
Ref. No. 005/UR/BMD/2015 FR.06 Laporan Program Uji Banding (Final Report) Ed./Rev./Date: 01/02/16-04-2015
www.ujiprofisiensi.com
Page 3 of 17
BMD Laboratory–Provider of Proficiency Testing Program Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email:
[email protected]
No.
Nitrat (NO3-)
Sulfat (SO42-)
Klordia (Cl-)
mg.L-1
mg.L-1
mg.L-1
BMD-206-15
12.10
5.96
7.40
Metode
Spectroquant
Spektrofotometri
Spectroquant
BMD-207-15
11.04
3.96
5.94
Metode
Spektrofotometri
Spektrofotometri
Titrimetri
BMD-208-15
2.02
3.77 ± 0.29
10.00 ± 0.24
Metode
SNI 06-2480-1991
SNI 6989.20:2009
SNI 6989.19:2009
BMD-209-15
10.21 ± 1.021
3.08 ± 0.30
7.09 ± 0.7
Metode
Ion Kromatografi
Ion Kromatografi
Ion Kromatografi
BMD-210-15
3.19
11.75
7.35
Metode
Spektrofotometri
Spektrofotometri
Titrimetri
BMD-213-15
9.81 ± 0.09
4.60 ± 0.45
8.01 ± 0.69
Metode
SNI 01-3554-2006
SNI 6989.20-2009
SNI 6989.19-2009
BMD-215-15
2.26 ± 0.07
3.20 ± 0.63
9.41 ± 0.44
Metode
SNI 01-3554-2006 (2.8)
SNI 01-3554-2006 (2.11)
SNI 01-3554-2006 (2.12)
Kode Lab.
5.
6.
7.
8.
9.
10.
11.
IV.1 Parameter Nitrat (NO3-) Sebanyak 10 (sepuluh) laporan hasil pengujian untuk parameter nitrat (NO3-) diterima oleh BMD Laboratory dan 5 (lima) diantaranya melaporkan nilai estimasi ketidakpastian diperluas dengan k=2 pada tingkat kepercayaan 95%. Dari hasil pengolahan data secara statistika diperoleh sebanyak 5 (lima) laboratorium masuk dalam kategori ‘memuaskan’ (|zscore|≤2,0), 2 (dua) laboratorium masuk dalam kategori ‘meragukan’ (2,0<|z-score|<3,0), dan 3 (tiga) laboratorium masuk dalam kategori ‘kurang memuaskan’ (|z-score|≥3,0). Hasil pengolahan data disajikan pada Tabel 3 dan Gambar 1.
Ref. No. 005/UR/BMD/2015 FR.06 Laporan Program Uji Banding (Final Report) Ed./Rev./Date: 01/02/16-04-2015
www.ujiprofisiensi.com
Page 4 of 17
BMD Laboratory–Provider of Proficiency Testing Program Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email:
[email protected]
Tabel 3. Hasil Pengolahan Data Parameter Nitrat (NO3-) No.
Kode Peserta
1. 2. 3. 4. 5. 6. 7. 8. 9. 10.
BMD-201-15 BMD-204-15 BMD-205-15 BMD-206-15 BMD-207-15 BMD-208-15 BMD-209-15 BMD-210-15 BMD-213-15 BMD-215-15
Nilai Laporan Peserta (mg/L) 9.84 3.46 2.44 12.10 11.04 2.02 10.21 3.19 9.81 2.26
U, k=2 (mg/L) 0.06 0.20 1.021 0.09 0.07
Nilai Evaluasi Uji Banding (mg/L) 8.85 8.85 8.85 8.85 8.85 8.85 8.85 8.85 8.85 8.85
SDPA (mg/L) 1.95 1.95 1.95 1.95 1.95 1.95 1.95 1.95 1.95 1.95
z-score
Kategori
0.51 -2.77 -3.29 1.67 1.12 -3.51 0.70 -2.91 0.49 -3.39
OK $ $$ OK OK $$ OK $ OK $$
Keterangan: Nilai evaluasi uji banding merupakan nilai konsensus (consensus value) laboratorium peserta yang diperoleh dari pengolahan data secara statistika menggunakan Robust Statistic Algoritma A Mean sesuai ISO 13528:2005(E) dengan sebelumnya dilakukan analisis menggunakan perangkat lunak Distribution AnalyzerTM (lihat Anex C). SDPA adalah simpangan baku yang ditetapkan sebesar 22% dari nilai evaluasi uji banding (assigned value), merupakan rata-rata dari simpangan baku Horwitz (SDHorwitz) dan simpangan baku Algoritma A Mean. U yaitu ketidakpastian diperluas dengan faktor cakupan (k)=2 pada tingkat kepercayaan 95%.
Ref. No. 005/UR/BMD/2015 FR.06 Laporan Program Uji Banding (Final Report) Ed./Rev./Date: 01/02/16-04-2015
www.ujiprofisiensi.com
Page 5 of 17
BMD Laboratory–Provider of Proficiency Testing Program
22.00 21.00 20.00 19.00 18.00 17.00 16.00 15.00 14.00 13.00 12.00 11.00 10.00 9.00 8.00 7.00 6.00 5.00 4.00 3.00 2.00 1.00 0.00
Cons. value = 8.85 mg/L Algoritma A Mean ISO 13528:2005(E) SDPA = 1.95 mg/L (22% Assigned Value)
+3SDPA +2SDPA
Cons. value
-2SDPA
BMD-206-15
BMD-207-15
BMD-209-15
BMD-201-15
BMD-213-15
BMD-204-15
BMD-210-15
BMD-205-15
BMD-215-15
-3SDPA
BMD-208-15
NO3- (mg/L)
Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email:
[email protected]
Lab. Code
©BMD Laboratory
Gambar 1. Grafik Hasil Uji Banding Parameter Nitrat (NO3-) IV.2 Parameter Sulfat (SO42-) Sebanyak 11 (sebelas) laporan hasil pengujian untuk parameter sulfat (SO42-) diterima oleh BMD Laboratory dan 7 (tujuh) diantaranya melaporkan nilai estimasi ketidakpastian diperluas dengan k=2 pada tingkat kepercayaan 95%. Dari hasil pengolahan data secara statistika diperoleh sebanyak 7 (tujuh) laboratorium masuk dalam kategori ‘memuaskan’ (|z-score|≤2,0), 2 (dua) laboratorium masuk dalam kategori ‘meragukan’ (2,0<|zscore|<3,0), dan 2 (dua) laboratorium masuk dalam kategori ‘kurang memuaskan’ (|zscore|≥3,0). Hasil pengolahan data disajikan pada Tabel 4 dan Gambar 2.
Ref. No. 005/UR/BMD/2015 FR.06 Laporan Program Uji Banding (Final Report) Ed./Rev./Date: 01/02/16-04-2015
www.ujiprofisiensi.com
Page 6 of 17
BMD Laboratory–Provider of Proficiency Testing Program Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email:
[email protected]
Tabel 4. Hasil Pengolahan Data Parameter Sulfat (SO42-) No.
Kode Peserta
1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11.
BMD-201-15 BMD-202-15 BMD-204-15 BMD-205-15 BMD-206-15 BMD-207-15 BMD-208-15 BMD-209-15 BMD-210-15 BMD-213-15 BMD-215-15
Nilai Laporan Peserta (mg/L) 4.68 5.64 3.89 4.50 5.96 3.96 3.77 3.08 11.75 4.60 3.20
U, k=2 (mg/L) 0.43 0.44 2.43 0.29 0.30 0.45 0.63
Nilai Evaluasi Uji Banding (mg/L) 4.23 4.23 4.23 4.23 4.23 4.23 4.23 4.23 4.23 4.23 4.23
SDPA (mg/L) 0.54 0.54 0.54 0.54 0.54 0.54 0.54 0.54 0.54 0.54 0.54
z-score
Kategori
0.82 2.58 -0.63 0.49 3.17 -0.50 -0.85 -2.11 13.80 0.68 -1.89
OK $ OK OK $$ OK OK $ $$ OK OK
Keterangan: Nilai evaluasi uji banding merupakan nilai konsensus (consensus value) laboratorium peserta yang diperoleh dari pengolahan data secara statistika menggunakan Robust Statistic Algoritma A Mean sesuai ISO 13528:2005(E) dengan sebelumnya dilakukan analisis menggunakan perangkat lunak Distribution AnalyzerTM (lihat Anex C), sedangkan SDPA adalah simpangan baku Horwitz (SDHorwitz). U yaitu ketidakpastian diperluas dengan faktor cakupan (k)=2 pada tingkat kepercayaan 95%.
Ref. No. 005/UR/BMD/2015 FR.06 Laporan Program Uji Banding (Final Report) Ed./Rev./Date: 01/02/16-04-2015
www.ujiprofisiensi.com
Page 7 of 17
BMD Laboratory–Provider of Proficiency Testing Program Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email:
[email protected]
13.00 Cons. value = 4.23 mg/L Algoritma A Mean ISO 13528:2005(E) SDPA = 0.54 mg/L (Horwitz Function)
12.00 11.00
SO42- (mg/L)
10.00 9.00 8.00 7.00 6.00
+3SDPA +2SDPA
5.00
Cons. value
4.00
-2SDPA -3SDPA
3.00 2.00 1.00
BMD-210-15
BMD-206-15
BMD-202-15
BMD-201-15
BMD-213-15
BMD-205-15
BMD-207-15
BMD-204-15
BMD-208-15
BMD-215-15
BMD-209-15
0.00
Lab. Code
©BMD Laboratory
Gambar 2. Grafik Hasil Uji Banding Parameter Sulfat (SO42-) IV.3 Parameter Klorida (Cl-) Sebanyak 11 (sebelas) laporan hasil pengujian untuk parameter klorida (Cl-) diterima oleh BMD Laboratory dan 7 (tujuh) diantaranya melaporkan nilai estimasi ketidakpastian diperluas dengan k=2 pada tingkat kepercayaan 95%. Dari hasil pengolahan data secara statistika diperoleh sebanyak 7 (tujuh) laboratorium masuk dalam kategori ‘memuaskan’ (|z-score|≤2,0), 2 (dua) laboratorium masuk dalam kategori ‘meragukan’ (2,0<|zscore|<3,0), dan 2 (dua) laboratorium masuk dalam kategori ‘kurang memuaskan’ (|zscore|≥3,0). Hasil pengolahan data disajikan pada Tabel 5 dan Gambar 3.
Ref. No. 005/UR/BMD/2015 FR.06 Laporan Program Uji Banding (Final Report) Ed./Rev./Date: 01/02/16-04-2015
www.ujiprofisiensi.com
Page 8 of 17
BMD Laboratory–Provider of Proficiency Testing Program Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email:
[email protected]
Tabel. 5 Hasil Pengolahan Data Parameter Klorida (Cl-) No.
Kode Peserta
1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11.
BMD-201-15 BMD-202-15 BMD-204-15 BMD-205-15 BMD-206-15 BMD-207-15 BMD-208-15 BMD-209-15 BMD-210-15 BMD-213-15 BMD-215-15
Nilai Laporan Peserta (mg/L) 7.12 16.19 6.92 11.41 7.40 5.94 10.00 7.09 7.35 8.01 9.41
U, k=2 (mg/L) 0.81 0.24 0.22 0.24 0.7 0.69 0.44
Nilai Evaluasi Uji Banding (mg/L) 7.58 7.58 7.58 7.58 7.58 7.58 7.58 7.58 7.58 7.58 7.58
SDPA (mg/L) 0.89 0.89 0.89 0.89 0.89 0.89 0.89 0.89 0.89 0.89 0.89
z-score
Kategori
-0.52 9.63 -0.74 4.28 -0.20 -1.84 2.70 -0.55 -0.26 0.48 2.04
OK $$ OK $$ OK OK $ OK OK OK $
Keterangan: Nilai evaluasi uji banding merupakan nilai konsensus (consensus value) laboratorium peserta yang diperoleh dari pengolahan data secara statistika menggunakan Robust Statistic Algoritma A Mean sesuai ISO 13528:2005(E) dengan sebelumnya dilakukan analisis menggunakan perangkat lunak Distribution AnalyzerTM (lihat Anex C), sedangkan SDPA adalah simpangan baku Horwitz (SDHorwitz). U yaitu ketidakpastian diperluas dengan faktor cakupan (k)=2 pada tingkat kepercayaan 95%.
Ref. No. 005/UR/BMD/2015 FR.06 Laporan Program Uji Banding (Final Report) Ed./Rev./Date: 01/02/16-04-2015
www.ujiprofisiensi.com
Page 9 of 17
BMD Laboratory–Provider of Proficiency Testing Program Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email:
[email protected]
18.00 17.00
Cons. value = 7.58 mg/L Algoritma A Mean ISO 13528:2005(E) SDPA = 0.89 mg/L (Horwitz Function)
16.00 15.00
Cl- (mg/L)
14.00 13.00 12.00 11.00
+3SDPA
10.00
+2SDPA
9.00 8.00
Cons. value
7.00 6.00
-2SDPA
5.00
-3SDPA
4.00 3.00
BMD-202-15
BMD-205-15
BMD-208-15
BMD-215-15
BMD-213-15
BMD-206-15
BMD-210-15
BMD-201-15
BMD-209-15
BMD-204-15
BMD-207-15
2.00
Lab. Code
©BMD Laboratory -
Gambar 3. Grafik Hasil Uji Banding Parameter Klorida (Cl )
V. PENUTUP Program uji banding antar laboratorium merupakan salah satu sarana jaminan mutu eksternal bagi laboratorium pengujian dan kalibrasi. Hasil dari program uji banding ini diharapkan dapat digunakan sebagai sarana evaluasi bagi laboratorium untuk meningkatkan kinerja dalam kegiatan pengujian yang dilakukan.
Ref. No. 005/UR/BMD/2015 FR.06 Laporan Program Uji Banding (Final Report) Ed./Rev./Date: 01/02/16-04-2015
www.ujiprofisiensi.com
Page 10 of 17
BMD Laboratory–Provider of Proficiency Testing Program Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email:
[email protected]
VI. REFERENSI [1] SNI 01-3553-2006 [2] SNI ISO/IEC 17025:2008 [3] ISO 13528:2005(E) [4] Distribution AnalyzerTM, Copyright © by Taylor Enterprises, Inc. USA [5] M. Thompson (2004) The Amazing Horwitz Function, AMC Technical Brief No.17, Royal Society of Chemistry
Ref. No. 005/UR/BMD/2015 FR.06 Laporan Program Uji Banding (Final Report) Ed./Rev./Date: 01/02/16-04-2015
www.ujiprofisiensi.com
Page 11 of 17
BMD Laboratory–Provider of Proficiency Testing Program Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email:
[email protected]
Annex A Rekapitulasi Hasil z-score Program Uji Banding No.
z-score
Kode Peserta Nitrat (NO3-)
Sulfat (SO42-)
Klorida (Cl-)
1.
BMD-201-15
0.51
OK
0.82
OK
-0.52
OK
2.
BMD-202-15
-
-
2.58
$
9.63
$$
3.
BMD-204-15
-2.77
$
-0.63
OK
-0.74
OK
4.
BMD-205-15
-3.29
$$
0.49
OK
4.28
$$
5.
BMD-206-15
1.67
OK
3.17
$$
-0.20
OK
6.
BMD-207-15
1.12
OK
-0.50
OK
-1.84
OK
7.
BMD-208-15
-3.51
$$
-0.85
OK
2.70
$
8.
BMD-209-15
0.70
OK
-2.11
$
-0.55
OK
9.
BMD-210-15
-2.91
$
13.80
$$
-0.26
OK
10.
BMD-213-15
0.49
OK
0.68
OK
0.48
OK
11.
BMD-215-15
-3.39
$$
-1.89
OK
2.04
$
Keterangan: OK : Kategori ‘memuaskan’ $ : Kategori ‘meragukan’ $$ : Kategori ‘kurang memuaskan’
Ref. No. 005/UR/BMD/2015 FR.06 Laporan Program Uji Banding (Final Report) Ed./Rev./Date: 01/02/16-04-2015
www.ujiprofisiensi.com
Page 12 of 17
BMD Laboratory–Provider of Proficiency Testing Program Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email:
[email protected]
Annex B Rekapitulasi Bias Nilai Peserta Terhadap Nilai Konsensus
Parameter Nitrat (NO3-) 4.00
b (mg/L)
2.00 0.00 -2.00 -4.00 -6.00 -8.00
Lab. Code
b (mg/L)
Parameter Sulfat(SO42-) 8.00 7.00 6.00 5.00 4.00 3.00 2.00 1.00 0.00 -1.00 -2.00
Lab. Code
Ref. No. 005/UR/BMD/2015 FR.06 Laporan Program Uji Banding (Final Report) Ed./Rev./Date: 01/02/16-04-2015
www.ujiprofisiensi.com
Page 13 of 17
BMD Laboratory–Provider of Proficiency Testing Program Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email:
[email protected]
Parameter Klorida (Cl-) 10.00
b (mg/L)
8.00 6.00 4.00 2.00 0.00 -2.00 -4.00
Lab. Code
Ref. No. 005/UR/BMD/2015 FR.06 Laporan Program Uji Banding (Final Report) Ed./Rev./Date: 01/02/16-04-2015
www.ujiprofisiensi.com
Page 14 of 17
BMD Laboratory–Provider of Proficiency Testing Program Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email:
[email protected]
Annex C Evaluasi Distribusi Normal Data Program Uji Banding No
Kode Peserta
Nilai Laporan Peserta (mg/L)
1 2 3 4 5 6 7 8 9 10
BMD-201-15 BMD-204-15 BMD-205-15 BMD-206-15 BMD-207-15 BMD-208-15 BMD-209-15 BMD-210-15 BMD-213-15 BMD-215-15
9.84 3.46 2.44 12.10 11.04 2.02 10.21 3.19 9.81 2.26
Sample Size = Average = Standard Deviation = Skewness = Excess Kurtosis =
10 6.637 4.248 0.05 -2.32
Test of Fit: p-value = (SK All) Decision = (SK Spec) Decision =
0.0033 Fail No specs
No
Kode Peserta
Nilai Laporan Peserta (mg/L)
1 2 3 4 5 7 8 9
BMD-201-15 BMD-204-15 BMD-205-15 BMD-206-15 BMD-207-15 BMD-209-15 BMD-210-15 BMD-213-15
9.84 3.46 2.44 12.10 11.04 10.21 3.19 9.81
Sample Size = Average = Standard Deviation = Skewness = Excess Kurtosis =
8 7.761 3.996 -0.53 -2.05
Test of Fit: p-value = (SK All) Decision = (SK Spec) Decision =
0.9604 Pass No specs
Normal Distribution Analysis (Parameter NO3-)
2.0200 7.0600 12.1000
Powered by: Distribution AnalyzerTM ©Taylor Enterprises, Inc. 5510 Fairmont Rd., Suite A Libertyville, IL 60048 USA www.variation.com
Normal Distribution Analysis (Parameter NO3-) BMD-208-15; BMD-215-15 were rejected
2.4400
7.2700
12.1000
With 95% confidence more than 99% of the values are between 2.4248 and infinity
Powered by: Distribution AnalyzerTM ©Taylor Enterprises, Inc. 5510 Fairmont Rd., Suite A Libertyville, IL 60048 USA www.variation.com
Ref. No. 005/UR/BMD/2015 FR.06 Laporan Program Uji Banding (Final Report) Ed./Rev./Date: 01/02/16-04-2015
www.ujiprofisiensi.com
Page 15 of 17
BMD Laboratory–Provider of Proficiency Testing Program Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email:
[email protected]
No
Kode Peserta
Nilai Laporan Peserta (mg/L)
1 2 3 4 5 6 7 8 9 10 11
BMD-201-15 BMD-202-15 BMD-204-15 BMD-205-15 BMD-206-15 BMD-207-15 BMD-208-15 BMD-209-15 BMD-210-15 BMD-213-15 BMD-215-15
4.68 5.64 3.89 4.50 5.96 3.96 3.77 3.08 11.75 4.60 3.20
Sample Size = Average = Standard Deviation = Skewness = Excess Kurtosis =
11 5.003 2.412 2.53 7.23
Test of Fit: p-value = (SK All) Decision = (SK Spec) Decision =
0.0034 Fail No specs
No
Kode Peserta
Nilai Laporan Peserta (mg/L)
1 2 3 4 5 6 7 8 10 11
BMD-201-15 BMD-202-15 BMD-204-15 BMD-205-15 BMD-206-15 BMD-207-15 BMD-208-15 BMD-209-15 BMD-213-15 BMD-215-15
4.68 5.64 3.89 4.50 5.96 3.96 3.77 3.08 4.60 3.20
Sample Size = Average = Standard Deviation = Skewness = Excess Kurtosis =
10 4.328 0.948 0.49 -0.47
Test of Fit: p-value = (SK All) Decision = (SK Spec) Decision =
0.7439 Pass No specs
Normal Distribution Analysis (Parameter SO42-)
3.0800
7.4150
11.7500
Powered by: Distribution AnalyzerTM ©Taylor Enterprises, Inc. 5510 Fairmont Rd., Suite A Libertyville, IL 60048 USA www.variation.com
Normal Distribution Analysis (Parameter SO42-)
BMD-210-15 was rejected
3.0800
4.5200
5.9600
With 95% confidence more than 99% of the values are between 0.1275 and 8.5285
Powered by: Distribution AnalyzerTM ©Taylor Enterprises, Inc. 5510 Fairmont Rd., Suite A Libertyville, IL 60048 USA www.variation.com
Ref. No. 005/UR/BMD/2015 FR.06 Laporan Program Uji Banding (Final Report) Ed./Rev./Date: 01/02/16-04-2015
www.ujiprofisiensi.com
Page 16 of 17
BMD Laboratory–Provider of Proficiency Testing Program Head Office: BMD Building Centre Jl. Puspiptek Raya No. 74B RT.01 RW.01 (Viktor) Serpong, Tangerang Selatan 15316 Tlp: 021-7563091 Fax: 021-7563291 Mobile: 0813 1516 2491 Email:
[email protected]
No
Kode Peserta
Nilai Laporan Peserta (mg/L)
1 2 3 4 5 6 7 8 9 10 11
BMD-201-15 BMD-202-15 BMD-204-15 BMD-205-15 BMD-206-15 BMD-207-15 BMD-208-15 BMD-209-15 BMD-210-15 BMD-213-15 BMD-215-15
7.12 16.19 6.92 11.41 7.40 5.94 10.00 7.09 7.35 8.01 9.41
Sample Size = Average = Standard Deviation = Skewness = Excess Kurtosis =
11 8.804 2.922 1.84 3.70
Test of Fit: p-value = (SK All) Decision = (SK Spec) Decision =
0.0085 Fail No specs
No
Kode Peserta
Nilai Laporan Peserta (mg/L)
1 3 4 5 6 7 8 9 10 11
BMD-201-15 BMD-204-15 BMD-205-15 BMD-206-15 BMD-207-15 BMD-208-15 BMD-209-15 BMD-210-15 BMD-213-15 BMD-215-15
7.12 6.92 11.41 7.40 5.94 10.00 7.09 7.35 8.01 9.41
Sample Size = Average = Standard Deviation = Skewness = Excess Kurtosis =
10 8.065 1.679 0.99 0.24
Test of Fit: p-value = (SK All) Decision = (SK Spec) Decision =
0.2731 Pass No specs
Normal Distribution Analysis (Parameter Cl-)
5.9400
11.0650
16.1900
Powered by: Distribution AnalyzerTM ©Taylor Enterprises, Inc. 5510 Fairmont Rd., Suite A Libertyville, IL 60048 USA www.variation.com
Normal Distribution Analysis (Parameter Cl-) BMD-202-15 was rejected
5.9400
8.6750
11.4100
With 95% confidence more than 99% of the values are between 0.6223 and 15.5077
Powered by: Distribution AnalyzerTM ©Taylor Enterprises, Inc. 5510 Fairmont Rd., Suite A Libertyville, IL 60048 USA www.variation.com
Ref. No. 005/UR/BMD/2015 FR.06 Laporan Program Uji Banding (Final Report) Ed./Rev./Date: 01/02/16-04-2015
www.ujiprofisiensi.com
Page 17 of 17
©BMD Laboratory 2015