Attila L Tóth CSc senior research fellow
tothal @mfa.kfki.hu
Attila L Tóth CSc senior research fellow
Research Institute for Technical Physics of the Hungarian Academy of Sciences (MTA MFKI) Budapest (Ujpest)
tothal @mufi.hu
MTA MFKI
Attila L Tóth CSc senior research fellow
Research Institute for Technical Physics and Materials Science of the Hungarian Academy of Sciences (MTA MFA) Budapest (Csillebérc)
tothal @mfa.kfki.hu
MTA MFA
Attila L Tóth CSc senior research fellow
Institute of Technical Physics and Materials Science, Research Centre for Natural Sciences, Hungarian Academy of Sciences (MTA TTK MFA) Budapest (Lágymányos)
toth.attila.lajos @ttk.mta.hu
MTA TTK MFA
SCANNING
1x1 um
10x10 cm 100.000x
SCANNING
1x1 um
10x10 cm 100.000x
MICROSCOPE A SEM mint analitikai mérőrendszer
Minden analitikai mérőrendszer (AMR) egyszerű elemekből épül fel. A mérő egység ( ME, a „tulajdonképpeni” analizátor) a vizsgált mintáról analitikai jelet szolgáltat, amiből az értelmező egység (ÉE) számolja ki a analitikai információt
A mérő egységben a vizsgálandó mintát egy δ reagenssel hozzuk köcsönhatásba (gerjesztés), mely a minta χ anyagtulajdonságainak függvényében kelti az η analitikai jelet
A SEM mint analitikai mérőrendszer
A SEM mint analitikai mérőrendszer
A SEM mint analitikai mérőrendszer
Analitikai jel (ÜZEMMÓD)
Felbontás laterális
mélységi
====================================================================
Szekunder elektron (SE)
1-10 nm
1-10 nm
E: jó felbontás (=sugárátmérõ), topográfiai kontraszt H: komplex jelképzés (pl.BE hozzájárulás) ---------------------------------------------------------------------------------------------------------------------
Visszaszórt elektron (BE)
0.1-0.5 um
0.1-0.5 um
E: rendszámkontraszt H: gyengébb felbontás (< sugárátmérõ) ---------------------------------------------------------------------------------------------------------------------
Áram (EBIC)
50-500 nm
0.1-1 um
E: Rácshibák és p-n átmenetek megfigyelhetõk H: Felületi rekombináció hatása ---------------------------------------------------------------------------------------------------------------------
Fény (CL)
10nm-1 um
0.2-1 um
E: fluoreszcens fázisok szelektiv leképzése H: kis intenzitás, sugárzási károsoodás halványit ---------------------------------------------------------------------------------------------------------------------
Röntgen sugár (EDS)
0.1-0.5 um
0.1-0.5 um
E: szimultán detektálás,nagy tömegérzékenység H: rossz felbontás, nagy holtidõ
(WDS)
0.2-1 um
0.1-1 um
E:jó felbontás, nagy intenzitás H: elemenkénti detektálás, rigorozus fókuszálásAES ---------------------------------------------------------------------------------------------------------------------
Auger el. (AES, Eo=2-5 keV)
10-100 nm
0.2-2 nm
E: Felületi monorétegek elemösszetétele H: Ultratiszta felület és ultrannagy vákuum ====================================================================
Analitikai jel (ÜZEMMÓD)
Felbontás laterális
mélységi
====================================================================
Szekunder elektron (SE)
1-10 nm
1-10 nm
E: jó felbontás (=sugárátmérõ), topográfiai kontraszt H: komplex jelképzés (pl.BE hozzájárulás) ---------------------------------------------------------------------------------------------------------------------
Szürkehályog operációkra használatos szikék élességének kvantitativ vizsgálata SEI ← Eo=25keV Eo=1,5keV ↓
SEI ↑ Eo=25keV BEI →
S2E-100s S2e-100c
Az élesség kvantitativ mérése
BEI-super TOPO üzemmód automatikus él-detektálással
Analitikai jel (ÜZEMMÓD)
Felbontás laterális
mélységi
====================================================================
Áram (EBIC)
50-500 nm
0.1-1 um
E: Rácshibák és p-n átmenetek megfigyelhetõk H: Felületi rekombináció hatása ---------------------------------------------------------------------------------------------------------------------
EBIC measurement techniques:
Beam induced currents and voltages (EBIC & EBIV) depending on the input impedance of measuring amplifier
Different (planar and X-sectional) EBIC sample preparation schemes of p/n junction (a,b) and Schottky (c.d) devices. The batteries should be used only in β conductivity measurement of insulators.
CC.mechanism Modulation
BARRIER electron voltaic effect (built in field of a p/n junction or a Schottky barrier)
Properties studied
EBIC
EBIC(x,y)
EBIC(Eo)
EBIC(t)
EBIC(V,x,y)
EBIC(T,x,y)
BULK
EBIC(t,T)
electron voltaic effect (internal field due to Fermi level variation)
EBIC(x,y,t,T)
EBIV(x,y)
β conductivity (external field)
SAMPLE CURRENT (absorbed electrons)
EBIV(x,y)
EBIC(x,y)
REBIC(x,y)
AEI(x,y)
electron/hole generation yield and energy diffusion length recombination velocity (electr. active) Xtal defects lifetime micro plasma effects recombination levels deep levels deep level distribution barrier height distribution of contaminants dopant distribution mobility (MOS) filed strength (MOS) local resistivity topography, mean Z
Recombination contrast of defects EBIC / SEI
Recombination activity of surface and volume defects as a function of temperature in InP LED
300 K
180 K
Recombination contrast of defects EBIC / SEI
Misfit network and threading dislocations in Si/Ge multilayer structure
FEG-SEM --- > ELECTRON ?
A LEO 1540XB cross beam ( Ga+ & e ) microscope and preparation system
[email protected]
Elektronoptikai lencsehibák
Nano-range: FEG-SEM • Gemini column:
– Schottky field emission gun – Beam booster – Special beam path -Short, shielded - Cross over free (Boersch eff.)
– Twin (Gemini) objective
- Electrostatic & electromagnetic - Low chromatic aberration even at low Eo
– In-lens SE detector
- sensitive for non-morphological SE contrast components , as work functions, surface potentials … - sensitive at short working distances (WD < 4 mm)
GEMINI FEG SEM basics
(courtesy LEO)
1500XB CrossBeam® with GEMINI column – low beam noise < 1 % – cross over free beam path , no significant Boersch effect, high depth of field – highly stable thermal FEG < 0.2 % /h variation – superb image resolution fhroughout the complete beam energy range, particularly down to 100 eV. – high resistance to ambient magnetic stray fields – constant conditions at sample surface eliminates ion-beam shift
RESULT:
nanoscope: excited volume in the nanometer range
GEMINI FEG SEM basics
(courtesy LEO)
Electron optics: „achromat” obj. Lens & InLens detector
GEMINI FEG SEM basics
(courtesy LEO)
Hol tartunk mi ? Szigeteló Al2O3 SEI képe 1.000.000x nagyításban
Részecske - optikai lencsehibák
Részecske - optikai lencsehibák
Ionoptikai lencsehibák
Ga ions : FIB
XB = FEG-SEM + FIB
FIB képalkotás: orientációs kontraszt
de …
[email protected]
Maris !
A FIB maratott acél minta SEM-SEI képe (45o elektronbeesési szög)
TEM-preparált acél minta FIB-SEI képe (88o ionbeesési szög)
[email protected]
Mar is !
A FIB maratott acél minta SEM-SEI képe (45o elektronbeesési szög)
TEM-preparált acél minta FIB-SEI képe (88o ionbeesési szög)
[email protected]
Ionoptikai lencsehibák
He ions : µscope
He ion - forrás
He ion mikroszkóp oszlop
Felbontás (SEI):
Felbontás (SEI):
Kontrasztmechanizmusok:
… és nem utolsósorban a STAGE
Massachusetts Institute of Technology, Hewlett Packard, and the Swiss Federal Laboratories for Material Science and Technology (EMPA) chose Raith ELPHY MultiBeam: Enables multitechnique helium ion microscope-based nanopatterning research
Xe ions : FIB
Xe ion - forrás
Xe ions : FIB .
Xe ions : FIB
Köszönettel tartozom: - Németh-Horváth Helga FEI , Prof. X-ray and EO Kft. - Dénes Éva, ISD DUNAFERR - Dr Vezendy László Nagykanizsai Kórház - Peter Gnauck, Carl Zeiss SMT
és köszönettel tartozom Önöknek a megtisztelÖ figyelemért