ADDENDUM DOKUMEN PENGADAAN ALAT GAS CHROMATOGRAFI (GC) Nomor: 06/Pokja/03/2013 Tanggal: 15 Maret 2013
ATAS
DOKUMEN PENGADAAN Nomor: 02/Pokja/03/2013 Tanggal: 13 Maret 2013
untuk Pengadaan ALAT GAS CHROMATOGRAFI (GC)
Kelompok Kerja Unit Layanan Pengadaan BADAN KETAHANAN PANGAN DAN PENYULUHAN ACEH Tahun Anggaran: 2013
Berdasarkan Berita Acara Penjelasan (Aanwijzing) pada hari jum’at tangggal lima belas bulan Maret tahun dua ribu tiga belas maka Dokumen Pengadaan Alat Gas Cromatografi (GC) Nomor 02/Pokja/3/2013 tanggal 13 Maret 2013, dilakukan perubahan sebagai berikut : 1. Spesifikasi Teknis SEMULA No.
NAMA DAN SPESIFIKASI PERALATAN
1 GAS CHROMATOGRAPH (FTD, FPD, ECD WITH AUTO INJECTOR & SAMPLER) Specifications: Column Oven Temperature range Dimensions Oven volume Temperature accuracy Temperature deviation Temperature variation coefficient Temperature program steps Programming rate setting range Total for all steps
: : : : : : : : :
Vol
1 Unit
+4°C to 450°C (using liquid Co2 gas: -50°C to 450°C) 280(W) x 280(H) x 175(D) mm 13.7 L Set value (K) ±1% (calibration at 0.01°C) <2°C max. (On 200 mm dia. Circumference 30mm from rear) <0.01°C/°C Up to 20 (cooling program possible) - 250 to 250°C/min 9999.99 minutes max.
Split/Splitless Injection Unit (SPL-2010 Plus) - Temperature range : Room temperature +5°C o 450°C - Flow Control Unit : Advanced Flow Controller (AFC) - Split/Splitless Mode : Pressure setting range: 0 to 970 kPa Programmable steps: 7 (pressure program possible). Programmed rate setting range: -400 to 400 kPa/min. Split ratio setting range: 0 to 9999,9. Total flow rate setting range: 0 to 1200 mL/min.
2 Set
Detectors Four detectors can be installed simulatenously and individually temperature controlled. Detector gas is electronically controlled by APC (Advance Pressure Control). The quantity that can be mounted simultaneously depends on the detector type. - Flame Thermionic Detector (FTD) 1 Set : To 450°C Temperature range : N 0.1 pgN/s (azobenzene) P 0.03 pgP/s (malathion) Minimum detection quantity Dynamic range : N, P 10 3 - Flame Photometric Detector (FPD) 1 Set : to 350°C. Temperature range : P 55 fgP/s (tributyl phosphate) S 3 pgS/s (dodecane thiol) Minimum detection quantity Dynamic range : Phosphorus mode: 104 Sulfur mode: 103 - Electron Capture Detector (ECD) 1 Set : to 350°C. Temperature range : 6 fg/s (Y-BHC) Minimum detection quantity Dynamic range : 10 4 Auto Injector for Liquid and Head Space, AOC-5000 Automatic injector for Liquid and Head Space Injection, with 2mL and 10/20mL vial tray, consisting of: 1pc PAL 80 X/Y unit 1pc Combi injection unit 1pc Control terminal & cable 1pc Power supply, Input Voltage 90-264V, 120/150W & cable 1pc Incubator fo in liquid mode Injection speed Syringe cleaning in headspace mode Injection speed Syringe cleaning Heated syringe Incubator oven Incubation temperature Agitation Incubation time Computer and Printer With Workstation GCSolution Software
: Selectable from 0.01µl/sec, up to 250 µl/sec Vial size: up to 600 1ml micro vials, 294 2ml standard vials, 96 10ml or 20ml vials, 4 deepwell microplates, 6 standard microplates. : Wash Station for 2 different solvents : Selectable from 0.01µl/sec, up to 5ml/sec Vial size: up to 294 2ml standard vials, 96 10ml or 20ml vials. : Wash Station for 2 different solvents : 30°C-150°C selectable in 1°C increments : 6 heated vial positions for 2ml/10ml/20ml vials : 30°C-200°C selectable in 1°C increments : Interval shaking 250rpm-750rpm : Up to 999 minutes selectable in 1 second in increments Minimum Intel Core i3 Processor, 2GB 1x2G CMel/CTino D3-1333, Windows 7 Professional 32-bit, HDD 1 Set 500GB 7200RPMSATA, Intel Graphics Media Accelerator (GMA) X4500HD, HP USB Optical Mouse, USB Standard Value Keyboard, HP 15-in-1 Media Card Reader, HP SuperMu Workstation Software 32-bit application (long-filename compatible), graphical user interface (Assistant Bar, etc.). Data Acquisition. 4ms minimum sampling time, snapshot function, supports single analysis and batch analysis, Batch Table Wizard, add or ins Data Processing and Data Analysis. Peak integration manipulation, identification (supports multiple relative retention times and grouping), quantitation (percentage area method, corrected percentage area method, internal standard method, external standard
Completed with: FTD collector (as consumable) FTD Collector Reactivation Kit Adaptor for FTD Collector Reactivation Kit S filter for FPD-2010 Plus P filter for FPD-2010 Plus Sn filter for FPD-2010 Plus ECD-2010 Detector Unit Carrier Gas Supply pipe 5 m Gas filter, molecular sieve He Gas + bottle Pressure Regulator He H2 Gas Supply Pipe (reverse threaded) 5 m H2 Gas + bottle Pressure Regulator H2 N2 Gas + bottle Presure Regulator N2 for ECD Oxygen Trap Oil-less air compressor Regulator filter Air filter for air compressor (silica gel) Joint Adapter Air supply pipe, 2 pcs 10-µL microsyringe Capillary Column, 30m, 0.25mm ID, 0.25 um thickness RS-232 Cable, 3 meter AUX Temperature Control Unit Tools Kit UPS 5 KVA Mounting hardware and interconnection cable Cycle Composer for AOC-5000 control AOC-5000 controlling software through GCMSsolution magnetic cap PTFE white with septum (100/ pkg) 10 mL screw vial (100 pcs/ pkg) 20 mL screw vial (100 pcs/ pkg) Magnet screw cap PTFE white for HSS & SPME (100 set/ pkg) Crimper for Headspace vial 20 mm Decapper for Headspace vial Liquid syringe 10µl for COMBI PAL and GC1 PAL Headspace Syringe HD-type 2.5ml for COMBI PAL
1 1 1 1 1 1 1 3 2 1 1 1 1 1 1 1 1 1 1 5 1 2 5 6 1 1 1 1 1 1 1 3 3 3 3 1 1 5 5
Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set
MENJADI No.
NAMA DAN SPESIFIKASI PERALATAN
1 GAS CHROMATOGRAPH (FTD, FPD, ECD WITH AUTO INJECTOR & SAMPLER) Specifications: Column Oven Temperature range Dimensions Oven volume Temperature accuracy Temperature deviation Temperature variation coefficient Temperature program steps Programming rate setting range Total for all steps
: : : : : : : : :
Split/Splitless Injection Unit (SPL-2010 - Temperature range : - Flow Control Unit : - Split/Splitless Mode :
Vol 1 Unit
+2°C to 460°C (using liquid Co2 gas: -50°C to 450°C) 275-285(W) x 275-285(H) x 165-180(D) mm 13 - 14 L Set value (K) ±1% (calibration at 0.01°C) <2°C max. (On 200 mm dia. Circumference 30mm from rear) <0.01°C/°C Up to 20 (cooling program possible) - 250 to 250°C/min 9999.99 minutes max. Plus) Room temperature +5°C o 450°C Advanced Flow Controller (AFC) Pressure setting range: 0 to 970 kPa Programmable steps: 7 (pressure program possible). Programmed rate setting range: -400 to 400 kPa/min. Split ratio setting range: 0 to 9999,9. Total flow rate setting range: 0 to 1200 mL/min.
2 Set
Detectors Four detectors can be installed simulatenously and individually temperature controlled. Detector gas is electronically controlled by APC (Advance Pressure Control). The quantity that can be mounted simultaneously depends on the detector type. - Flame Thermionic Detector (FTD) 1 Set : To 450°C Temperature range : N 0.1 pgN/s (azobenzene) P 0.03 pgP/s (malathion) Minimum detection quantity Dynamic range : N, P 10 3 - Flame Photometric Detector (FPD) 1 Set : to 325°C - 400°C. Temperature range : P 55 fgP/s (tributyl phosphate) S 3 pgS/s (dodecane thiol) Minimum detection quantity Dynamic range : Phosphorus mode: 104 Sulfur mode: 103 - Electron Capture Detector (ECD) 1 Set : to 325°C - 400°C. Temperature range : 6 fg/s (Y -BHC) Minimum detection quantity Dynamic range : 10 4 Auto Injector for Liquid and Head Space, AOC-5000 Automatic injector for Liquid and Head Space Injection, with 2mL and 10/20mL vial tray, consisting of: 1pc PAL 80 X/Y unit 1pc Combi injection unit 1pc Control terminal & cable 1pc Power supply, Input Voltage 90-264V, 120/150W & cable 1pc Incubator fo in liquid mode Injection speed Syringe cleaning in headspace mode Injection speed Syringe cleaning Heated syringe Incubator oven Incubation temperature Agitation Incubation time Computer and Printer With Workstation GCSolution Software
: Selectable from 0.01µl/sec, up to 250 µl/sec Vial size: up to 600 1ml micro vials, 294 2ml standard vials, 96 10ml or 20ml vials, 4 deepwell standard microplates. : microplates, Wash Station6for 2 different solvents : Selectable from 0.01µl/sec, up to 5ml/sec Vial size: up to 294 2ml standard vials, 96 10ml or 20ml vials. : Wash Station for 2 different solvents : 30°C-150°C selectable in 1°C increments : 6 heated vial positions for 2ml/10ml/20ml vials : 30°C-200°C selectable in 1°C increments : Interval shaking 250rpm-750rpm : Up to 999 minutes selectable in 1 second in increments Minimum Intel Core i3 Processor, 2GB 1x2G CMel/CTino D3-1333, Windows 7 Professional 32-bit, 1 Set HDD 500GB 7200RPMSATA, Intel Graphics Media Accelerator (GMA) X4500HD, HP USB Optical Mouse, USB Standard Value Keyboard, HP 15-in-1 Media Card Reader, HP SuperMu Workstation Software 32-bit application (long-filename compatible), graphical user interface (Assistant Bar, etc.). Data Acquisition. 4ms minimum sampling time, snapshot function, supports single analysis and batch analysis, Batch Table Wizard, add or ins Data Processing and Data Analysis. Peak integration manipulation, identification (supports multiple relative retention times and grouping), quantitation (percentage area method, corrected percentage area method, internal standard method, external standard
Completed with: FTD collector (as consumable) FTD Collector Reactivation Kit Adaptor for FTD Collector Reactivation Kit S filter for FPD-2010 Plus P filter for FPD-2010 Plus Sn filter for FPD-2010 Plus ECD-2010 Detector Unit Carrier Gas Supply pipe 5 m Gas filter, molecular sieve He Gas + bottle Pressure Regulator He H2 Gas Supply Pipe (reverse threaded) 5 m H2 Gas + bottle Pressure Regulator H2 N2 Gas + bottle Presure Regulator N2 for ECD Oxygen Trap Oil-less air compressor Regulator filter Air filter for air compressor (silica gel) Joint Adapter Air supply pipe, 2 pcs 10-µL microsyringe Capillary Column, 30m, 0.25mm ID, 0.25 um thickness RS-232 Cable, 3 meter AUX Temperature Control Unit Tools Kit UPS 5 KVA Mounting hardware and interconnection cable Cycle Composer for AOC-5000 control AOC-5000 controlling software through GCMSsolution magnetic cap PTFE white with septum (100/ pkg) 10 mL screw vial (100 pcs/ pkg) 20 mL screw vial (100 pcs/ pkg) Magnet screw cap PTFE white for HSS & SPME (100 set/ pkg) Crimper for Headspace vial 20 mm Decapper for Headspace vial Liquid syringe 10µl for COMBI PAL and GC1 PAL Headspace Syringe HD-type 2.5ml for COMBI PAL
1 1 1 1 1 1 1 3 2 1 1 1 1 1 1 1 1 1 1 5 1 2 5 6 1 1 1 1 1 1 1 3 3 3 3 1 1 5 5
Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set Set
2. Hal-hal lainnya mengacu pada Dokumen Pengadaan Alat Gas Cromatografi (GC) Nomor 02/Pokja/3/2013 tanggal 13 Maret 2013
3. Addendum Dokumen pengadaan ini merupakan bagian yang tidak terpisahkan dengan Dokumen Pengadaan Alat Gas Cromatografi (GC) Nomor 02/Pokja/3/2013 tanggal 13 Maret 2013 Demikian addendum dokumen pengadaan ini dibuat untuk dapat dipergunakan sebagaimana mestinya. Banda Aceh, 15 Maret 2013 Tertanda Pokja Pengadaan Barang/Jasa ULP Badan Ketahanan Pangan dan Penyuluhan Aceh