SOLAR I-V PV single phase certifier and I-V curve checker
Rel. 1.01 – 20/07/10 Pag 1 of 5
1. SOLAR I-V MAIN FEATURES SOLAR I-V: use as PV certifier SOLAR I-V performs all tests on Single phase PV plants by using of SOLAR-02 remote unit which, after a preliminary synchronisation, save in independent way the values of irradiance and temperature. Only at the end of test the remote unit should be connected via wireless RF with the master to download the recorded data
SOLAR I-V: use as I-V curve checker SOLAR I-V allows the on field measurement of I-V curve as well as of the main parameters of a single module and of a whole photovoltaic system up to a maximum of 1000V and 10A
With SOLAR-02 remote unit the irradiance and temperature measured values are shown at display also in independent mode (ideal solution during a pre-test on installation) besides test/recording with SOLAR I-V
The HT304 reference cell permits to performs solar irradiance measurements both on PV modules in Monocrystalline and Polycrystalline silicon material
SOLAR I-V
Rel. 1.01 – 20/07/10
PV single phase certifier and I-V curve checker
Pag 2 of 5
2. ELECTRICAL SPECIFICATIONS Accuracy is calculated as ± [% readings + (no. of digits) * resolution] at 23°C ± 5°C, con relative humidity <80%HR
2.1. CERTIFIER OF SINGLE PHASE PV INSTALLATION DC Voltage Range (V) 15.0 ÷ 999.9
Resolution (V) 0.1
Accuracy ± (0.5%rdg + 2dgt)
Resolution (V) 0.1
Accuracy ± (0.5%rdg + 2dgt)
AC TRMS Voltage Range (V) 50.0 ÷ 265.0 Max crest factor: 1.5
DC Current (by external transducer clamp) Range (mV) -1100 ÷ -5 5 ÷ 1100
Resolution (mV)
Accuracy
0.1
± (0.5%rdg + 0.6mV)
The value of current is ALWAYS displayed with positive sign ; The value of current transduced in voltage less then 5mV is zeroed
AC TRMS Current (by external transducer clamp) Range (mV) 1 ÷ 1200
Resolution (mV) 0.1
Frequency (Hz) 47.5 ÷ 63.0
Accuracy ± (0.5%rdg + 0.6mV)
Max crest factor: 2.0 ; The value of current transduced in voltage less then 5mV is zeroed
FS DC & AC clamp (A)
Resolution (A)
1< FS ≤ 10 10< FS ≤ 100 100< FS ≤ 1000
0.001 0.01 0.1
Minimum read value (A) DC AC 0.05 0.01 0.5 0.1 5A 1
DC Power (Vmis > 150V, Imis > 10%FS) FS clamp (A) 1< FS ≤ 10 10< FS ≤ 100 100< FS ≤ 1000
Range (W]) 0.000k ÷ 9.999k 10.00k ÷ 99.99k 0.000k ÷ 9.999k 10.00k ÷ 99.99k 0.00k ÷ 99.99k 100.0k ÷ 999.9k
Resolution (W) 0.001k 0.01k 0.001k 0.01k 0.01k 0.1k
Accuracy
± (0.7%rdg)
Vmis = voltage correspondent to measured power
AC Single phase power (@ PF = 1, Vmis > 200V, Imis > Imis > 10%FS) FS clamp (A) 1< FS ≤ 10 10< FS ≤ 100 100< FS ≤ 1000
Range (W) 0.000k ÷ 9.999k 10.00k ÷ 99.99k 0.000k ÷ 9.999k 10.00k ÷ 99.99k 0.00k ÷ 99.99k 100.0k ÷ 999.9k
Vmis = voltage correspondent to measured power
Resolution (W) 0.001k 0.01k 0.001k 0.01k 0.01k 0.1k
Accuracy
± (0.7%rdg)
SOLAR I-V
Rel. 1.01 – 20/07/10
PV single phase certifier and I-V curve checker
Frequency Range (Hz) 47.5 ÷ 63.0Hz
Resolution (Hz) 0.1
Accuracy ±(0.2%rdg+1dgt)
Irradiance (by reference cell HT304) Range (mV) 1.0 ÷ 100.0
Resolution (mV) 0.1
Accuracy ±(1.0%rdg + 5dgt)
Temperature (by external probe PT300N) Range (°C) -20.0 ÷ 100.0
Resolution (°C) 0.1
Accuracy ± (1.0%rdg +1°C)
Pag 3 of 5
SOLAR I-V
Rel. 1.01 – 20/07/10
PV single phase certifier and I-V curve checker
Pag 4 of 5
2.2. I-V CURVE MEASUREMENT VDC Voltage Range (V) 5.0 ÷ 999.9
Resolution (V) 0.1
Accuracy ±(1.0%rdg+2dgt)
(*) The I-V curve and Rs measurements start for VDC > 15V and the accuracy is defined for VDC > 20V
IDC Current (by internal sensor) – Detection of I-V Curve Range (A) 0.10 ÷ 10.00
Resolution (A) 0.01
Accuracy ±(1.0%rdg+2dgt)
Max Power (@ Vmpp >30V, Impp >2A) Range (W) 50 ÷ 999
Resolution (W) 1
Accuracy ±(1.0%rdg+6dgt)
Vmpp = voltage on point of maximum power ; Impp = current on point of maximum power NOTE 1 Max measurable value of Power must include FF value(~ 0.7) Æ Pmax = 1000V x 10A x 0.7 = 7000W NOTE 2Test is stopped and the message “Thermal instability” occurs if the instrument detects Voltage > 700V and Current I >3A, I > 0.038*V + 37.24 - 0.5
Irradiance (by reference cell HT304) Range (mV) 1.0 ÷ 100.0
Resolution (mV) 0.1
Accuracy ±(1.0%rdg+5dgt)
Temperature of module (by PT300N probe) Range (°C) -20.0 ÷ 100.0
Resolution (°C) 0.1
Accuracy ±(1.0%rdg+1°C)
SOLAR I-V PV single phase certifier and I-V curve checker
Rel. 1.01 – 20/07/10 Pag 5 of 5
3. GENERAL SPECIFICATIONS DISPLAY AND MEMORY: Features: Memory capacity: Saved data: POWER SUPPLY: SOLAR I-V internal power supply: Autonomy of SOLAR I-V:
128x128pxl custom LCD with backlight 256kbytes max 99 yield test ; >249 curves (I-V curve test)
SOLAR-02 power supply: SOLAR-02 max recording time (@ IP=5s):
6x1.5V alkaline batteries type LR6, AA, AM3, MN 1500 > 200 curve (measurement of I-V curve) approx 120 hours (yield test) 4x1.5V alkaline batteries type AAA LR03 approx 1.5h
OUTPUT INTERFACE PC communication port: Interface with SOLAR-02 :
optical/USB wireless RF comunication (max distance 1m)
MECHANICAL FEATURES Size: Weight (batteries included):
235 (L) x 165 (W) x 75 (D) mm 1.2kg
ENVIRONMENTAL CONDITIONS: Reference temperature: Working temperature: Working humidity: Storage temperature (batt. not included): Storage humidity:
23°C ± 5°C 0° ÷ 40°C < 80% UR -10 ÷ 60°C < 80% UR
GENERAL REFERENCE STANDARDS: Safety: Safety of measurement accessories: I-V curve measurement: Insulation: Pollution degree: Overvoltage category: Max altitude of use:
IEC/EN61010-1 IEC/EN61010-031 IEC/EN60891 double insulation 2 CAT II 1000V DC, CAT III 300V AC to ground Max 1000V among inputs P1, P2, C1, c2 2000m
This instrument complies with the requirements of the European Low Voltage Directives 2006/95/EEC (LVD) and EMC 2004/108/EEC
EURO-INDEX Service Onderhoud en kalibratie van meetinstrumenten ....................................................................... De Nederlandse vestiging van EURO-INDEX beschikt over een bijzonder modern service- en kalibratielaboratorium. Hier worden de meetinstrumenten uit het assortiment preventief onderhouden, gerepareerd, gekalibreerd en indien nodig gejusteerd. Het service- en kalibratielaboratorium van EURO-INDEX is verdeeld in verschillende disciplines, gebaseerd op het soort meetinstrument en de gemeten grootheden. Druk
Elektrische
Gasanalyse
Gasdetectie
Temperatuur (inclusief
Luchtsnelheid
infrarood temperatuurmeting en thermografie)
grootheden en luchthoeveelheid
Waarom een kalibratiecertificaat? .................................................................................................. Een kalibratiecertificaat vermeldt hoeveel een meetinstrument afwijkt ten opzichte van onze, naar (inter)nationale standaarden herleidbare, kalibratiemiddelen. Bij de meetresultaten op het certificaat wordt tevens vermeld of het meetinstrument voldoet aan de specificaties die door de fabrikant zijn opgegeven. Zonder kalibratiecertificaat kunt u er vanuit gaan dat de meter voldoet aan de fabrieksspecificaties, maar aantonen kunt u dit niet. Een testcertificaat van de fabrikant is te beknopt om de lineairiteit aan te tonen en is niet geregistreerd op naam (wat wel degelijk een vereiste is).
EC
IM
EN
De
SP
SP
EC
IM EN
KWS® ................................................................................................................................................... KWS® is een uniek servicesysteem voor uw meetinstrumenten met periodiek onderhoud en kalibratie. Veel zaken worden voor u geregeld, zodat u zonder zorgen gebruik kunt maken van uw meetinstrumenten. prijs staat vast voor de levensduur van het instrument (mits de KWS® behandeling volgens herkalibratieadvies periodiek wordt uitgevoerd in het EURO-INDEX kalibratielaboratorium) Geen arbeidsloon bij de KWS® behandeling Kalibratie voor justage (voorkalibratie) indien mogelijk Indien nodig justage en (na)kalibratie
Reparatie
en preventief onderhoud oproep met het advies voor herkalibratie Controle op functionaliteit van het instrument Vijf jaar historie voor alle gegevens 10% korting op onderdelen Serienummerregistratie Franco retourlevering Gratis
RvA accreditatie ............................................................................................................................ Het kalibratielaboratorium van EURO-INDEX beschikt sinds 21 augustus 1997 over een RvA accreditatie naar NEN-EN-ISO/IEC 17025. Deze accreditatie geldt voor verschillende grootheden, zoals gespecificeerd in de scope bij accreditatienummer K105 op www.rva.nl. Test- en meetinstrumenten voor grootheden die deel uitmaken van de gespecificeerde scope, kunnen worden voorzien van een RvA kalibratiecertificaat. De metingen worden uitgevoerd met standaarden waarvan de herleidbaarheid naar (inter)nationale standaarden, ten overstaan van de Raad voor Accreditatie, is aangetoond. In het Multilateral Agreement zijn de meeste Europese landen overeengekomen elkaars accreditaties te accepteren. Hierdoor is een RvA kalibratiecertificaat internationaal geaccepteerd. Bovendien wordt op een RvA kalibratiecertificaat de meetonzekerheid van de gerapporteerde meetresultaten vermeld. Verhuur van meetinstrumenten ........................................................................................................... EURO-INDEX biedt een assortiment meetinstrumenten te huur aan. Na deskundig advies van onze productspecialisten, wordt bepaald welk instrument u nodig heeft voor uw specifieke werkzaamheden. De instrumenten worden compleet met accessoires geleverd, inclusief herleidbaar kalibratiecertificaat. Wijzigingen voorbehouden EURO-INDEX® VL 13001
www.rva.nl www.rva.nl www.rva.nl www.rva.nl www.rva.nl www.rva.nl www.rva.nl www.rva.nl www.rva.nl www.rva.nl www.rva.nl www.rva.nl www.rva.nl www.rva.nl www.rva.nl www.rva.nl www.rva.nl www.rva.nl www.rva.nl www.rva.nl www.rva.nl www.rva.nl www.rva.nl www.rva.nl